Medline medical convenience kits labeled as recallZ-2788-2026 · Medline Industries, LP · initiated Jun 11, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medline Industries, LP is issuing a recall for specific Medline convenience kits containing BD PurPrep 26 mL which may be contaminated with low levels of Bacillus species.
FDA root cause category: Process change control
What the company is doing
Medline Industries, LP issued an URGENT MEDICAL DEVICE RECALL notice to its consignees via email and USPS first class mail. The notice explained the issue with the device, potential harms, and requested the following: REQUIRED ACTION: 1. Immediately check stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete the response form. Please list the quantity of affected product in inventory on the form. Even if there is no affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com 3. Upon receipt of the response form, the consignee account will receive over-labels to place on affected inventory, with instructions for the staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. Distributors and those who transferred the affected product were directed to notify those to which the product was distributed or transferred. For contact the Recall Department at 866-359-1704 or [email protected].
Recall details
- Recall number
- Z-2788-2026
- Status
- Open, Classified
- Initiated by the company
- Jun 11, 2026
- Posted by the FDA
- Jul 23, 2026
- Recalling firm
- Medline Industries, LP
- Quantity
- 155 kits
- Product code
- OEZ
- Distribution
- US: CT, FL, KS, OH, OR, NY, MD, MT, VA, MA.
Product description
Medline medical convenience kits labeled as: 1) KIT PEDIATRIC OPEN HEART MOSES, Medline Kit SKU DYNJ910103B; 2) BASIC HEART CDS, Medline Kit SKU DYNJ910961A.
Code information (models, lots, serial numbers)
Medline Kit SKU DYNJ910103B, UDI/DI 10198459346873 (each) 40198459346874 (case), Lot Number 25FBR237; Medline Kit SKU DYNJ910961A, UDI/DI 10198459213342 (each) 40198459213343 (case), Lot Number 25ABM761; Medline Kit SKU DYNJ910961A, UDI/DI 10198459213342 (each) 40198459213343 (case), Lot Number 25BBD107; Medline Kit SKU DYNJ910961A, UDI/DI 10198459213342 (each) 40198459213343 (case), Lot Number 25FBD534; Medline Kit SKU DYNJ910961A, UDI/DI 10198459213342 (each) 40198459213343 (case), Lot Number 25FBL772.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYNJ910961AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99329).
- Z-2789-2026 · Open, ClassifiedMedline medical convenience kits labeled as: 1) SHOULDER PACK, Medline Kit SKU DYNJ39590O; 2) SHOULDER PACK, Medline Kit SKU DYNJ88566; 3) MAJOR ORTHO KIT, Medline Kit SKU DYNJ…
- Z-2790-2026 · Open, ClassifiedMedline medical convenience kits labeled as: NEURO PACK, Medline Kit SKU DYNJ59397P
- Z-2791-2026 · Open, ClassifiedMedline medical convenience kits labeled as: 1) TOTAL JOINT PACK Medline Kit SKU DYNJ65065A: 2) EXTREMITY/FOOT PACK Medline Kit SKU DYNJ65066B: 3) ANTERIOR CERVICAL-SMH Medline…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page