Photera 600, phototherapy device recallZ-2743-2026 · Daavlin Distributing Company · initiated Mar 5, 2026

Class II Open, Classified All recalls posted in July 2026

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Daavlin Distributing Company or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).

FDA root cause category: Software Design Change

What the company is doing

Daavlin Distributing Company initially notified its consignees about the issue beginning on 3/5/2026 via telephone following with by an email. The firm a follow up URGENT MEDICAL DEVICE Advisory notice on 4/9/2026 with the same information via US mail and email. On 7/10/2026 the firm sent an URGENT MEDICAL DEVICE RECALL notice to all non-responding consignees. The notice explained the issue, potential risk to health, and requested the following: "" Verify your device is operating ClearLink Controller Software Version 3.17 as follows: a) Ensure your machine is plugged in and on b) Locate and press the gear icon in the top right corner of the controller screen c) Select Machine Info d) Select Software Version e) Review Software Information " Confirm the serial number of your device is 123567890abc. The serial number can be found in all the following places: " For devices that remain in the original packaging, it can be found on the outside of the box " For devices that have been removed from the original packaging, it can be found on the back of your machine centered on the top of the Spine Cover or on the bottom of the Spine Cover "Once you confirm your device is identified as an affected device operating ClearLink Controller Software Version 3.17, contact Daavlin Technical Services immediately (see contact information below). Daavlin will place the device into Fixed Calibration Mode as an interim mitigation until ClearLink Controller Software Version 3.18 can be installed. " Complete the enclosed Recall Response Form and return it within 10 business days. Share this notice with all users of this device. " If you distributed the affected device to another facility, immediately provide them with a copy of this notice. " For questions, please contact: Daavlin Technical Services, Phone: (419) 633-3415, Email: [email protected]

Recall details

Recall number
Z-2743-2026
Recall class
Class II
Status
Open, Classified
Initiated by the company
Mar 5, 2026
Posted by the FDA
Jul 15, 2026
Recalling firm
Daavlin Distributing Company
Quantity
26 units
Product code
FTC
FDA submissions
K212510
Distribution
US Nationwide distribution.

Product description

Photera 600, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders

Code information (models, lots, serial numbers)

Serial Numbers: 6-09033NB4CL, 6-09037NB4CL, 6-09031NB4CL, 6-09538NB4CL, 6-09038NB4CL, 6-09039NB4CL, 6-09040NB4CL, 6-09041NB4CL, 6-09035NB4CL, 6-09034NB4CL, 6-09030NB4CL, 6-09026NB4CL, 6-09023NB4CL, 6-09028NB4CL, 6-09025NB4CL, 6-09043NB4CL, 6-09029NB4CL, 6-09044NB4CL, 6-09022NB4CL, 6-09027NB4CL, 6-09036NB4CL, 6-09045NB4CL, 6-09024NB4CL, 6-07827NB4CL, 6-09539NB4CL, 6-09042NB4CL.

Other devices under the same FDA submission

1 device cleared under K212510 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.