Skip to content

Rapid Refill Continuous Injection System, UPN M00566001 recallZ-2725-2026 · Boston Scientific Corporation · initiated Jun 30, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Boston Scientific Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Devices contain recalled Medline Namic Angiographic Control Syringe with Rotating Adaptor. There is a risk of syringe loosening or disconnecting from the manifold, resulting in loose connection or complete disconnection. May lead to difficulty connecting syringe, inability to aspirate or inject properly, fluid leakage, or separation of the syringe from the system, potentially requiring device exchange and prolonging the procedure.

FDA root cause category: Under Investigation by firm

What the company is doing

On June 30, 2026, the firm began notifying US customers via letters titled "Urgent Medical Device Removal - Immediate Action Required." Customers were instructed to immediately stop further use or distribution of affected product, and segregate it. Continue to follow the standard care when monitoring patients treated with the contact. Product is to be returned to Boston Scientific.

Recall details

Recall number
Z-2725-2026
Status
Open, Classified
Initiated by the company
Jun 30, 2026
Posted by the FDA
Jul 10, 2026
Recalling firm
Boston Scientific Corporation
Quantity
153,921 (US); 119,274 (OUS)
Product code
KYZ
Distribution
US Nationwide distribution. OUS consignee list pending.

Product description

Rapid Refill Continuous Injection System, UPN M00566001

Code information (models, lots, serial numbers)

Show the full code information (3k characters)

UDI-DI (GTIN) 08714729285151 Lots 33786611, 33825949, 33826026, 33949916, 33968571, 33968575, 34079981, 34153104, 34153413, 34153415, 34203075, 34203077, 34153411, 34203079, 34203115, 34386605, 34386609, 34386613, 34429656, 34429658, 34595249, 34693559, 34616613, 34616615, 34712433, 34763427, 34763967, 34805723, 34805725, 34805729, 35006995, 35006999, 35056793, 35208946, 35186845, 35208950, 35397314, 35208952, 35305436, 35498030, 35448316, 35498028, 35551523, 35626939, 35574463, 35688118, 35688460, 35828714,35831703, 35848483, 35877768, 35995282, 36049415, 36159194, 36219882, 36294100, 36303614, 36332230, 31177006, 31180206, 31666278, 31329922, 31356107, 31666276, 31798400, 31798133, 31798135, 31798137, 31852246, 32162450, 32208753, 32255526, 32255528, 32496142, 32661870, 32696046, 32696050, 32696052, 32752697, 32773682, 32791475, 32791477, 32820865, 32838612, 32791479, 33119798, 32977346, 32977348, 33030942, 33120320, 33158008, 33158006, 33354311, 33354315, 33354317, 33403248, 33403252, 33696882, 33696884, 33696886, 33696888, 33750883, 33750885, 33750887, 36332232, 36497262, 36159192, 36587704, 36612263, 36739668, 36785215, 36785217, 36814560, 36845486, 36983883, 36983886, 37291592, 37059471, 37417957, 37417959, 37429808, 37456936, 37472523, 37585745, 37627037, 37654892, 37941477, 37974508, 38033021, 38033023, 38124799, 38143635, 38165214, 38182214, 38201123, 37959187, 38353956, 38374833, 38430201, 33968573, 34386611, 32838610, 38205377, 38429054, 34066207, 34429654, 34616611, 34712429, 34763971, 34805727, 35006997, 35448312, 35448314, 35626937, 35688021, 35848481, 35687399, 31616842, 31601573, 31616848, 31616850, 31616846, 32208755, 32455154, 35877766, 36776858, 37037796, 37200706, 37276220, 37291594, 37639238, 38154387, 35305438, 31797614, 33354313, 33696890, 36470789, 37309672, 37396601, 37407323, 37472521, 35056797, 35626933, 33786613, 33825951, 33950422, 34965218, 35056795, 36101020, 36100418, 31356105, 32208759, 32661868, 33030940, 33679650, 37059473, 37041524, 37633157, 37991645, 38137850, 38393609, 33968569, 34703963, 36319481, 33649863, 37019542, 37493275, 34616609, 34386607, 34763969, 34805721, 35942747, 31852244, 32977350, 33354319, 37884043, 38001944, 32752695, 37189280, 37612931, 38703180, 34429700, 33968567, 34017321, 34568306, 34712431, 34873735, 34873737, 34965420, 35007081, 35186843, 35208948, 35397312, 35448310, 35305600, 35626935, 35778515, 35778517, 35778513, 35989120, 36027670, 36049413, 36303612, 36319483, 36312073, 36353951, 36437019, 36517931, 36409546, 36537384, 36622267, 36750117, 36750119, 36994750, 36962014, 37407325, 37456934, 37766480, 37660978, 37764898, 37723376, 37991643, 38856580, 38178214, 38337853, 38326961, 38410111, 38414293, 38576233, 38586789, 38622115, 38633220, 38676778, 38649491, 38680582, 38864691, 38687308, 38876174, 38901702, 31170922, 31601571, 31616844, 31852242, 31852240, 32162448, 32208757, 32696048, 32723872, 32723874, 32838608, 32964550, 32977344, 33119794, 33119796, 33257466, 33334705, 33403250, 33649865, 33933383

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page