Rapid Refill Continuous Injection System, UPN M00566001 recallZ-2725-2026 · Boston Scientific Corporation · initiated Jun 30, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Boston Scientific Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Devices contain recalled Medline Namic Angiographic Control Syringe with Rotating Adaptor. There is a risk of syringe loosening or disconnecting from the manifold, resulting in loose connection or complete disconnection. May lead to difficulty connecting syringe, inability to aspirate or inject properly, fluid leakage, or separation of the syringe from the system, potentially requiring device exchange and prolonging the procedure.
FDA root cause category: Under Investigation by firm
What the company is doing
On June 30, 2026, the firm began notifying US customers via letters titled "Urgent Medical Device Removal - Immediate Action Required." Customers were instructed to immediately stop further use or distribution of affected product, and segregate it. Continue to follow the standard care when monitoring patients treated with the contact. Product is to be returned to Boston Scientific.
Recall details
- Recall number
- Z-2725-2026
- Status
- Open, Classified
- Initiated by the company
- Jun 30, 2026
- Posted by the FDA
- Jul 10, 2026
- Recalling firm
- Boston Scientific Corporation
- Quantity
- 153,921 (US); 119,274 (OUS)
- Product code
- KYZ
- Distribution
- US Nationwide distribution. OUS consignee list pending.
Product description
Rapid Refill Continuous Injection System, UPN M00566001
Code information (models, lots, serial numbers)
Show the full code information (3k characters)
UDI-DI (GTIN) 08714729285151 Lots 33786611, 33825949, 33826026, 33949916, 33968571, 33968575, 34079981, 34153104, 34153413, 34153415, 34203075, 34203077, 34153411, 34203079, 34203115, 34386605, 34386609, 34386613, 34429656, 34429658, 34595249, 34693559, 34616613, 34616615, 34712433, 34763427, 34763967, 34805723, 34805725, 34805729, 35006995, 35006999, 35056793, 35208946, 35186845, 35208950, 35397314, 35208952, 35305436, 35498030, 35448316, 35498028, 35551523, 35626939, 35574463, 35688118, 35688460, 35828714,35831703, 35848483, 35877768, 35995282, 36049415, 36159194, 36219882, 36294100, 36303614, 36332230, 31177006, 31180206, 31666278, 31329922, 31356107, 31666276, 31798400, 31798133, 31798135, 31798137, 31852246, 32162450, 32208753, 32255526, 32255528, 32496142, 32661870, 32696046, 32696050, 32696052, 32752697, 32773682, 32791475, 32791477, 32820865, 32838612, 32791479, 33119798, 32977346, 32977348, 33030942, 33120320, 33158008, 33158006, 33354311, 33354315, 33354317, 33403248, 33403252, 33696882, 33696884, 33696886, 33696888, 33750883, 33750885, 33750887, 36332232, 36497262, 36159192, 36587704, 36612263, 36739668, 36785215, 36785217, 36814560, 36845486, 36983883, 36983886, 37291592, 37059471, 37417957, 37417959, 37429808, 37456936, 37472523, 37585745, 37627037, 37654892, 37941477, 37974508, 38033021, 38033023, 38124799, 38143635, 38165214, 38182214, 38201123, 37959187, 38353956, 38374833, 38430201, 33968573, 34386611, 32838610, 38205377, 38429054, 34066207, 34429654, 34616611, 34712429, 34763971, 34805727, 35006997, 35448312, 35448314, 35626937, 35688021, 35848481, 35687399, 31616842, 31601573, 31616848, 31616850, 31616846, 32208755, 32455154, 35877766, 36776858, 37037796, 37200706, 37276220, 37291594, 37639238, 38154387, 35305438, 31797614, 33354313, 33696890, 36470789, 37309672, 37396601, 37407323, 37472521, 35056797, 35626933, 33786613, 33825951, 33950422, 34965218, 35056795, 36101020, 36100418, 31356105, 32208759, 32661868, 33030940, 33679650, 37059473, 37041524, 37633157, 37991645, 38137850, 38393609, 33968569, 34703963, 36319481, 33649863, 37019542, 37493275, 34616609, 34386607, 34763969, 34805721, 35942747, 31852244, 32977350, 33354319, 37884043, 38001944, 32752695, 37189280, 37612931, 38703180, 34429700, 33968567, 34017321, 34568306, 34712431, 34873735, 34873737, 34965420, 35007081, 35186843, 35208948, 35397312, 35448310, 35305600, 35626935, 35778515, 35778517, 35778513, 35989120, 36027670, 36049413, 36303612, 36319483, 36312073, 36353951, 36437019, 36517931, 36409546, 36537384, 36622267, 36750117, 36750119, 36994750, 36962014, 37407325, 37456934, 37766480, 37660978, 37764898, 37723376, 37991643, 38856580, 38178214, 38337853, 38326961, 38410111, 38414293, 38576233, 38586789, 38622115, 38633220, 38676778, 38649491, 38680582, 38864691, 38687308, 38876174, 38901702, 31170922, 31601571, 31616844, 31852242, 31852240, 32162448, 32208757, 32696048, 32723872, 32723874, 32838608, 32964550, 32977344, 33119794, 33119796, 33257466, 33334705, 33403250, 33649865, 33933383
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Rapid Refill M00566000Boston Scientific Corporation · SYRINGE, IRRIGATING (NON DENTAL)1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.