Skip to content

5-PLASMA Bivalirudin AMR Verification Kit, Model Number… recallZ-2894-2026 · Aniara Diagnostica LLC · initiated Jul 1, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Aniara Diagnostica LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Stability issues with the reconstituted lyophilized human plasma and bivalirudin matrix have been identified. These stability issues lead to a degradation of measurable bivalirudin in the controls, calibrators, and linearity values.

FDA root cause category: Process control

What the company is doing

Email notifications were sent to customers on 6/17/26. Please reply to this email as acknowledgment you have received this initial notification and please include the quantity of each product which is remaining in your possession. A notification letter dated 7/1/26 was sent to customers. Actions to be implemented: " Destroy all stock of 5-PLASMA Bivalirudin Control, 5-PLASMA Bivalirudin Calibrator, & 5-PLASMA AMR Verification Kits - Please permanently deface the box labels and vial labels before disposal. " Complete and return the Awareness Acknowledgement form. " If product has been consumed please add brief statement to the Comment section, complete the End User Information, and return the Awareness Acknowledgement form.

Recall details

Recall number
Z-2894-2026
Status
Open, Classified
Initiated by the company
Jul 1, 2026
Posted by the FDA
Aug 5, 2026
Recalling firm
Aniara Diagnostica LLC
Quantity
4 units
Product code
JIT
Distribution
US Nationwide distribution in the states of IL, NC, TN.

Product description

5-PLASMA Bivalirudin AMR Verification Kit, Model Number 5D-45905

Code information (models, lots, serial numbers)

UDI-DI: 05070002420378, Lot Number: KE132603

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 99344).

Sources and updates