Skip to content

Fort Defiance Industries (FDI), Automated Field Steam… recallZ-2983-2026 · Fort Defiance Industries, LLC · initiated Jul 8, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Fort Defiance Industries, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.

FDA root cause category: Component design/selection

What the company is doing

Between the dates of July 8, 2026 - July 16, 2026 Fort Defiance Industries (FDI) issued a Urgent Medical Device Recall to affected consignees vi Email. FDI ask consignees to take the following actions: 1. Devices can continue in operation until the fix is implemented. 2. In the interim, users should avoid touching the WRS power connector during operation as a precaution. 3. Please contact FDI and order the upgrade kit complete with instructions to perform this Field Correction. 4. After receiving the kit, please review the instructions and determine if you are competent to perform the update. The instructions will state alternate options to complete the repair (i.e. send to military depot). 5. Perform the repair 6. FDI requests a phone call or e-mail confirmation when the upgrade is completed. This confirmation can be a phone call or e-mail message.

Recall details

Recall number
Z-2983-2026
Status
Open, Classified
Initiated by the company
Jul 8, 2026
Posted by the FDA
Aug 18, 2026
Recalling firm
Fort Defiance Industries, LLC
Product code
FLE
FDA submissions
K141009
Distribution
Worldwide - US Nationwide distribution in the states of MD, VA, NJ, TN and the countries of Canada, Lithuania, Spain.

Product description

Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131

Code information (models, lots, serial numbers)

Lot Code: All serial numbers/ GTIN: (01) 00862992000327

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates