Skip to content

Syva Emit tox Serum Tricyclic Antidepressant… recallZ-2881-2026 · Siemens Healthcare Diagnostics, Inc. · initiated Jun 26, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

out-of-specification sodium azide concentration in Syva Emit Tox Serum Tricyclic Antidepressant Calibrator/Controls Lot U2. While the assay results were not impacted, the overall health risk to lab technicians due to this issue is low, as exposure is limited by small volumes and a 0.2% post-reconstitution sodium azide concentration, with the worst outcome being mild skin or eye irritation and no serious harm anticipated.

FDA root cause category: Under Investigation by firm

What the company is doing

On June 26, 2026, the firm began notifying customers via Urgent Medical Device Correction letters. Customers were informed of the issue and provided with revised Safety Data Sheets (SDS) specific to the impacted lots for handling and disposal information.

Recall details

Recall number
Z-2881-2026
Status
Open, Classified
Initiated by the company
Jun 26, 2026
Posted by the FDA
Aug 4, 2026
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
494 US; 114 OUS
Product code
LFG
FDA submissions
K821743
Distribution
Worldwide - US Nationwide distribution in the states CA, ID, IL, IN, MA, MS, NC, NM, NY, OH, OK, OR, PA, WA, and Northern Mariana Islands part of USA and the countries of Canada, South Africa, Brazil, Chile, Colombia.

Product description

Syva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog No. 7C219UL

Code information (models, lots, serial numbers)

UDI-DI 00842768014321 Lot U2

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page