Atellica CH Urine Albumin recallZ-2890-2026 · Siemens Healthcare Diagnostics, Inc. · initiated Jul 6, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

processing Atellica CH Urine Albumin (UAlb) tests on Atellica CH and Atellica CI analyzers may lead to the following issues: Falsely depressed Atellica CH Urea Nitrogen (UN_c) results; Falsely depressed Atellica CH Enzymatic Hemoglobin A1c (A1c_E) results; and Increased rate of reaction cuvette QC failures and Delayed results due to QC failures.

FDA root cause category: Under Investigation by firm

What the company is doing

On July 6, 2026, Siemens Healthineers issued a Urgent Medical Device Recall to affected consignees via Fed Ex. Siemens Healthineers ask consignees to take the following actions: 1. Discontinue Atellica CH Urine Albumin (UAlb) testing and transition to the Atellica CH Microalbumin_2 (ALB_2) reagent, SMN 11097610, and Atellica CH Microalbumin_2 Calibrator (ALB_2 CAL), SMN 11099435, as an alternate test solution. 2. Replace all reaction cuvette segments following discontinuation of UAlb testing. Refer to the Atellica Solution Online Help for instructions on replacing reaction cuvette segments. 3. Complete the Field Correction Effectiveness Check Form attached to this letter and return within 30 days. 4.Credit will be issued for unexpired UAlb reagent (SMN 11537225), UAlb calibrator (SMN 11537247), and reaction cuvettes segments, as applicable. 5. Please review this letter with your Medical Director to determine the appropriate course of action, including whether review of any previously generated results is warranted. 6. Please retain this letter with your laboratory records and forward this letter to those who may have received the affected product.

Recall details

Recall number
Z-2890-2026
Status
Open, Classified
Initiated by the company
Jul 6, 2026
Posted by the FDA
Aug 5, 2026
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
7018 units
Product code
JIR
Distribution
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY and the countries of Australia, Austria, Bahamas, Belgium, Brazil, Canada, Chile, Cura¿ao, St. Eus, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Ireland, Israel,¿Italy,¿Latvia, Malaysia, Netherlands, New Zealand, Norway, Portugal, Republic of Korea, Romania, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, United¿Kingdom.¿

Product description

Atellica CH Urine Albumin REF: 11537225

Code information (models, lots, serial numbers)

All Lots within expiry/UDI: 00630414611099

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.