Dermalon recallZ-2920-2026 · Covidien, LP · initiated Jun 24, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Covidien, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Product may contain a perforation or tear in the sterile barrier packaging. This issue may result in biocontamination of the sterile sutures, which may lead to infection, loss of vision if in ophthalmic utilization, or prolonged surgery.
FDA root cause category: Equipment maintenance
What the company is doing
The firm began notifying consignees on 7/7/26 via email, mail, or electronically. Notices were titled, "URGENT: VOLUNTARY MEDICAL DEVICE RECALL." Customers were informed of the manufacturing nonconformance. If products in scope of this field action have been used in patients under the customer's care, the customer is instructed to monitor for signs of infection or vision issues due to the possibility of biocontamination. Customers were instructed to immediately identify, cease use, and quarantine all unused impacted product. Unused impacted product is to be returned to Medtronic. The recall notice should be passed on to all those who need to be aware within the customer's organization or to any organization where devices have been transferred.
Recall details
- Recall number
- Z-2920-2026
- Status
- Open, Classified
- Initiated by the company
- Jun 24, 2026
- Posted by the FDA
- Aug 7, 2026
- Recalling firm
- Covidien, LP
- Quantity
- 135,875 total
- Product code
- GAR
- FDA submissions
- K981582
- Distribution
- US Nationwide and the countries of Australia, Austria, Belgium, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Guadeloupe, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Kosovo, Lebanon, Lithuania, Luxembourg, Mexico, Namibia, Netherlands, North Macedonia, Northern Ireland, Norway, Oman, Panama, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan.
Product description
Dermalon 88861757-31 DERMALON* 4-0 BLU 75CM C13 88861757-31 88861799-31 DERMALON* 4-0 BLU 75CM C14 88861799-31
Code information (models, lots, serial numbers)
88861757-31 UDI-DI 10884521070912 Lot D6A3020FY UDI-DI 20884521070919 Lot D6A3020FY 88861799-31 UDI-DI 20884521071060 Lot D6A2861FY
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Dermalon 88861757-31Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall1 report
- Dermalon 88861799-31Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall1 report
Product lines
Other devices under the same FDA submission
47 devices cleared under K981582 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Surgilon 88861919-41Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE2 recalls2 reports
- Surgilon 88861919-31Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE2 recalls0 reports
- Surgilon 88861919-51Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE2 recalls0 reports
- Surgilon 88861919-71Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE2 recalls0 reports
- Dermalon 88861749-24Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall4 reports
- Dermalon 88861741-11Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall2 reports
- Dermalon 88861747-11Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall0 reports
- Dermalon 88861757-41Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall0 reports
- Surgilon 88861883-51Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall0 reports
- Surgilon 88861915-71Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall0 reports
- Surgilon 88861919-61Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall0 reports
- Surgilon 88861932-51Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99390).
- Z-2918-2026 · Open, ClassifiedV-Loc 90 Absorbable Wound Closure Device VLOC 90 0 VL 22 GS-21 VLOCM0336 VLOC 90 0 VL 22 GS-25 VLOCM0436 VLOCM0436 VLOCM0003 V-LOC* 90 4-0 CLR 15CM P12X12 VLOCM0003 VL…
- Z-2919-2026 · Open, ClassifiedV-Loc 180 Absorbable Wound Closure Device BOX BEDSIDE STAPLING IS KING BOX06614V4 BOX BY PASS FINAL BOX07405V1 BOX BYPASS CARDON BOX01056V2 BOX BYPASS DR LAGAE BOX0…
- Z-2921-2026 · Open, ClassifiedSurgipro SP-623 SURGIPRO* 2-0 BLU 75CM SC2 X36 SP-623
- Z-2922-2026 · Open, ClassifiedSurgipro II S8698GX SURGIPRO II BL 5/0 18 P-13 S8698GX
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page