Skip to content

GETINGE Y SENSOR Module, SERVO-U, SERVO-N, Model Number… recallZ-2963-2026 · Maquet Critical Care AB · initiated Jun 30, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Maquet Critical Care AB or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Maquet Critical Care AB, a subsidiary of Getinge, is voluntarily recalling Y sensor module, part of the SERVO-U/-N ventilator system, due to missing production safety electrical testing.

FDA root cause category: Under Investigation by firm

What the company is doing

Getinge issued an URGENT MEDICAL DEVICE CORRECTION notice was issued to its consignees on 7/9/2026 via FedEx. The notice explained the issue, potential risk to health, and requested the following: "Prior to the replacement, Getinge requests that you take the following actions: 1. Ensure all affected product in your inventory has been identified. 2. Please ensure this message is forwarded to any individuals that need notification within your organization or any organization where the affected devices have been transferred. 3. Contact Technical Service if you have questions at 888-943-8872 (888 9GETUSA), Option 4, Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone). 4. Complete and return the enclosed response form to Getinge by e-mailing a copy to [email protected]. For questions regarding this communication, please contact contact Technical Service if you have any questions at 888-943-8872 (888 9GETUSA), Option 4, Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).

Recall details

Recall number
Z-2963-2026
Status
Open, Classified
Initiated by the company
Jun 30, 2026
Posted by the FDA
Aug 14, 2026
Recalling firm
Maquet Critical Care AB
Quantity
9 units
Product code
CBK
FDA submissions
K201874
Distribution
US Nationwide distribution in the states of CA, GA, MN, NY.

Product description

GETINGE Y SENSOR Module, SERVO-U, SERVO-N, Model Number 6693662; ventilator, continuous, facility use

Code information (models, lots, serial numbers)

DUI/DI 07325710000304, Serial/Lot Numbers: 21049, 21088, 21089, 21093, 21100, 21103 21108, 21114, 21115.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

3 devices cleared under K201874 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates