Halyard LAVH PACK, Model/Catalog Number: USAR053-11 recallZ-2964-2026 · AVID Medical, Inc. · initiated Jul 10, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from AVID Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A customer complaint reported that the Clearify Visualization System (5303907) leaked in a kit. As a result of Avid s investigation, it was determined that the component was subjected to EO sterilization which is not in alignment with the manufacturer s recommendation.
FDA root cause category: Process control
What the company is doing
Owens & Minor issued an URGENT: MEDICAL DEVICE FIELD CORRECTION notice to its consignees on 7/10/2026 via email. The notice explained the issue, potential risks to health, and requested the following: "ACTIONS TO TAKE: " DO NOT USE AND DISCARD the Clearify Visualization System (Component# 5303907) located in the convenience kit identified in Table 1 above. All other components in the kits may continue to be used. " Add warning labels to affected AVID kits to ensure all users are aware of the actions to take prior to use. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. This label should be applied in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. Please contact your Sales Representative if pre-printed labels are needed. " Customers (end-users): If you have received product from a distributor, please contact the distributor or your sales representative with any questions you may have. " Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. " Complete the attached Recall Response Form and return by July 22, 2026."
Recall details
- Recall number
- Z-2964-2026
- Status
- Open, Classified
- Initiated by the company
- Jul 10, 2026
- Posted by the FDA
- Aug 14, 2026
- Recalling firm
- AVID Medical, Inc.
- Quantity
- 40 packs
- Product code
- OKV
- Distribution
- US Nationwide distribution in the state of TX.
Product description
Halyard LAVH PACK, Model/Catalog Number: USAR053-11
Code information (models, lots, serial numbers)
UDI/DI: 10809160372357, Lot No.: 1669869, Expiration Date: 4/30/2027
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Halyard USAR053-11_0004Avid Medical, Inc. · Obstetrical kit2 recalls0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.