FilmArray NGDS Warrior Panel REF NGDS-ASY-0007 recallZ-2877-2026 · Biofire Defense · initiated Jul 16, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Biofire Defense or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Firm has identified mismatches between the outbreak strain and the reference strain used in the development of its qualitative, multiplexed, nucleic acid-based in vitro diagnostic panel test.
FDA root cause category: Under Investigation by firm
What the company is doing
On 07/16/2026, an "ADVISORY NOTICE URGENT: MEDICAL DEVICE CORRECTION" Letter was sent via FedEx Priority Overnight to customer informing them that, due to the emergence of the 2026 Bundibugyo outbreak strain, BioFire Defense has conducted testing to evaluate performance of the Orthoebolavirus spp. assays on the FilmArray NGDS Warrior Panel. This preliminary testing has identified a potential reduction in sensitivity to the 2026 Bundibugyo outbreak strain which could result in Orthoebolavirus spp. Not Detected results when testing on the FilmArray NGDS Warrior Panel. Customers are instructed to: " Distribute this letter and the updated IFU (revision 5) to all appropriate personnel in your laboratory, retain a copy in your files, and forward this information to all parties that may use this product including others to whom you may have transferred FilmArray NGDS Warrior Panel product. Please follow applicable ITAR protocols for distribution. " Train and ensure awareness by laboratory personnel on the Additional Information and Required Actions included in the updated IFU (Table 6, Pg. 24). " Complete the Acknowledgement of Receipt. The web form can be found here: https://go.biofiredefense.com/FSCA-2601-Warrior-potential-ebola-not-detected-US or fill out the accompanying document. For Questions - contact BioFire Defense Customer Technical Support at [email protected] or via telephone at 1-801-262-3592.
Recall details
- Recall number
- Z-2877-2026
- Status
- Open, Classified
- Initiated by the company
- Jul 16, 2026
- Posted by the FDA
- Aug 4, 2026
- Recalling firm
- Biofire Defense
- Quantity
- 122 kits (3,660 pouches)
- Product code
- PRD
- FDA submissions
- K170883
- Distribution
- U.S. Nationwide distribution in the states of AL, CA, DC, FL, GA, HI, KY, LA, MA, MD, MN, NC, NH, NV, NY, OH, RI, TN, TX, UT, VA and WA.
Product description
FilmArray NGDS Warrior Panel REF NGDS-ASY-0007 UDI-DI code: (01)00851458005136 The FilmArray NGDS Warrior Panel is a qualitative, multiplexed, nucleic acid-based in vitro diagnostic test intended for use with the FilmArray 2.0 system. The FilmArray NGDS Warrior Panel detects and identifies Bacillus anthracis, Yersinia pestis, Francisella tularensis, Coxiella burnetii, Ebola virus, and Marburg virus nucleic acids directly from human whole blood (EDTA).
Code information (models, lots, serial numbers)
Part Number: NGDS-ASY-0007 UDI-DI code: (01)00851458005136 Lot Numbers: 311825D 312025D 313526D 322925D
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- FilmArray® NGDS Warrior Panel NGDS-ASY-0007Biofire Defense, LLC · Biothreat Microbial Agents Multiplex Nucleic Acid Detection System2 recalls0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page