Medline medical convenience kits, labeled as recallZ-2996-2026 · Medline Industries, LP · initiated Jul 8, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Specific item(s) and lot(s) of Medline convenience kits containing Pilling Wecksorb Neurosponges and Neurosponge Strips are being recalled due to potential out-of-specification levels of endotoxin of the neurosponges.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Medline Industries issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 7/8/2026 via USPS and email. The notice explained the issue, potential adverse health consequences, and requested the following: "REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-26-112-FGX2 Recall Code: 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product." For questions, contact the Recall Department at 866-359-1704 or [email protected].
Recall details
- Recall number
- Z-2996-2026
- Status
- Open, Classified
- Initiated by the company
- Jul 8, 2026
- Posted by the FDA
- Aug 21, 2026
- Recalling firm
- Medline Industries, LP
- Quantity
- 3468 kits
- Product code
- OJG
- Distribution
- US Nationwide distribution in the states of AZ, CA, IA, ID, IL, IN, MA, MD, MI, MN, MO, MT, NY, OH, TX, WI, WY.
Product description
Medline medical convenience kits, labeled as: 1) NEURO CDS, Medline Kit SKU CDS985246J; 2) NEURO PACK, Medline Kit SKU DYNJ32770V; 3) NEURO BUNDLE-LF, Medline Kit SKU DYNJ40095D; 4) NEURO PACK-LF, Medline Kit SKU DYNJ42844D; 5) NEURO PACK, Medline Kit SKU DYNJ45121I; 6) NEURO LAMINECTOMY PACK, Medline Kit SKU DYNJ58012A; 7) NEURO PACK SJ, Medline Kit SKU DYNJ63477A; 8) BASIC NEURO PACK, Medline Kit SKU DYNJ83071A.
Code information (models, lots, serial numbers)
Show the full code information (5k characters)
Medline Kit SKU CDS985246J, UDI/DI 10193489232394 (each) 40193489232395 (case), Lot Number 22HMB295; Medline Kit SKU CDS985246J, UDI/DI 10193489232394 (each) 40193489232395 (case), Lot Number 22IMB860; Medline Kit SKU DYNJ32770V, UDI/DI 10195327090111 (each) 40195327090112 (case), Lot Number 22FMB806; Medline Kit SKU DYNJ40095D, UDI/DI 10195327281250 (each) 40195327281251 (case), Lot Number 23GMI744; Medline Kit SKU DYNJ42844D, UDI/DI 10193489880564 (each) 40193489880565 (case), Lot Number 25EBV362; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 25CMH980; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 25FMB096; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 25GMA066; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 25GMH940; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 25BMI081; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 25IMC735; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 24DME519; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 24EMG256; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 24GMJ491; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 24IMA243; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 24JMA162; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 24JMD784; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 24KMI470; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 24ABN077; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 24ABN475; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 24DBA116; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 23AMH395; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 23EMA407; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 23HBH164; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 23HBO192; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 23HBV863; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 23IBF802; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 23IBS028; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 23JBV613; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 22BME230; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 22CMA406; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 22EMB248; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 22EMI134; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 22HME564; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 22IMA455; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 22JMI068; Medline Kit SKU DYNJ45121I, UDI/DI 10193489887631 (each) 40193489887632 (case), Lot Number 22LMB601; Medline Kit SKU DYNJ58012A, UDI/DI 10193489863994 (each) 40193489863995 (case), Lot Number 24BMI624; Medline Kit SKU DYNJ58012A, UDI/DI 10193489863994 (each) 40193489863995 (case), Lot Number 24EME751; Medline Kit SKU DYNJ58012A, UDI/DI 10193489863994 (each) 40193489863995 (case), Lot Number 23AMJ022; Medline Kit SKU DYNJ58012A, UDI/DI 10193489863994 (each) 40193489863995 (case), Lot Number 23CMB364; Medline Kit SKU DYNJ58012A, UDI/DI 10193489863994 (each) 40193489863995 (case), Lot Number 23EMD743; Medline Kit SKU DYNJ58012A, UDI/DI 10193489863994 (each) 40193489863995 (case), Lot Number 23FMH669; Medline Kit SKU DYNJ58012A, UDI/DI 10193489863994 (each) 40193489863995 (case), Lot Number 23JME474; Medline Kit SKU DYNJ58012A, UDI/DI 10193489863994 (each) 40193489863995 (case), Lot Number 22DMH314; Medline Kit SKU DYNJ58012A, UDI/DI 10193489863994 (each) 40193489863995 (case), Lot Number 22LME727; Medline Kit SKU DYNJ63477A, UDI/DI 10195327123451 (each) 40195327123452 (case), Lot Number 22EMB958; Medline Kit SKU DYNJ63477A, UDI/DI 10195327123451 (each) 40195327123452 (case), Lot Number 22EMC777; Medline Kit SKU DYNJ83071A, UDI/DI 10195327452490 (each) 40195327452491 (case), Lot Number 23HMF326.
Devices named in this recall
8 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYNJ42844DMedline Industries, Inc. · Neurological tray5 recalls0 reports
- Medline Industries, Inc. DYNJ40095DMedline Industries, Inc. · Neurological tray4 recalls0 reports
- Medline Industries, Inc. CDS985246JMedline Industries, Inc. · Neurological tray3 recalls0 reports
- Medline Industries, Inc. DYNJ45121IMedline Industries, Inc. · Neurological tray3 recalls0 reports
- Medline Industries, Inc. DYNJ63477AMedline Industries, Inc. · Neurological tray3 recalls0 reports
- Medline Industries, Inc. DYNJ83071AMedline Industries, Inc. · Neurological tray3 recalls0 reports
- Medline Industries, Inc. DYNJ58012AMedline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ32770VMedline Industries, Inc. · Neurological tray2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99489).
- Z-2994-2026 · Open, ClassifiedMedline medical convenience kits, labeled as: 1) SINUS PACK-LF, Medline Kit SKU DYNJ0966146F; 2) ENT PACK, Medline Kit SKU DYNJ40154D; 3) ENT PACK, Medline Kit SKU DYNJ40154F; …
- Z-2995-2026 · Open, ClassifiedMedline medical convenience kits, labeled as: 1) CRANIOTOMY PACK, Medline Kit SKU DYNJT6545; 2) ORTHO SPINE PACK, Medline Kit SKU DYNJ88056A; 3) ENT SINUS PACK, Medline Kit SKU…
- Z-2997-2026 · Open, ClassifiedMedline medical convenience kits, labeled as: 1) LAMINECTOMY PACK, Medline Kit SKU DYNJT6548; 2) ORTHO SPINE PACK, Medline Kit SKU DYNJ44504G; 3) NEURO SPINE PACK, Medline Kit …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page