Onli recallZ-2998-2026 · Hollister Ulc · initiated Jul 14, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Hollister Ulc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Hollister has identified that select lots of the catheters identified in the above table may exhibit a protrusion caused by a manufacturing defect which might lead to genitourinary trauma or bleeding.

FDA root cause category: Process control

What the company is doing

Hollister issued an URGENT PRODUCT FIELD SAFETY NOTICE to its consignees on 7/16/2026 via USPS and email. The notice explained the issue, risk to health, and requested the consignee inspect each catheter for the potential presence of a protrusion (defect) located approximately 2 cm from the funnel. If a protrusion is identified, the catheter should not be used. Users should also review and follow the Instructions for Use (IFU) and ensure that the device is opened, handled, and inserted in accordance with the IFU. The notice also requested the following actions be taken: "Please follow the steps below: 1. Please complete and return the response form attached as soon as possible, even if you do not have product in your possession and send the form to [email protected]. 2. If you have supplied or transferred any potentially affected product to another facility or organization, provide that facility with a copy of this letter immediately. For further information or to request replacement devices please contact us: By phone Please contact one of the following: If you are a consumer or healthcare professional, please call: 1.888.808.7456 If you are a dealer, distributor, Institution, or pharmacy, please call: 1.800.323.4060 option 1"

Recall details

Recall number
Z-2998-2026
Status
Open, Classified
Initiated by the company
Jul 14, 2026
Posted by the FDA
Aug 21, 2026
Recalling firm
Hollister Ulc
Quantity
2526120 units
Product code
GBM
FDA submissions
K211436
Distribution
US Nationwide distribution.

Product description

Onli 17cm Hydrophilic Urethral Intermittent Catheter, labeled as: 1) 82081 Ch08-17 Onli Hydrophilic Urethral Intermittent Catheter, SKU/REF 82081; 2) 82101 Ch10-17 Onli Hydrophilic Urethral Intermittent Catheter, SKU/REF 82101; 3) 82121 Ch12-17 Onli Hydrophilic Urethral Intermittent Catheter, SKU/REF 82121; 4) 82141 Ch14-17 Onli Hydrophilic Urethral Intermittent Catheter, SKU/REF 82141; 5) 82084 Ch08 40cm Onli Hydrophilic Urethral Intermittent Catheter, SKU/REF 82084; 6) 82104 Ch10 40cm Onli Hydrophilic Urethral Intermittent Catheter, SKU/REF 82104; 7) 82124 Ch12 40cm Onli Hydrophilic Urethral Intermittent Catheter, SKU/REF 82124; 8) 82144 Ch14-40cm Onli Hydrophilic Urethral Intermittent Catheter, SKU/REF 82144; 9) 82164 Ch16 40cm Onli Hydrophilic Urethral Intermittent Catheter, SKU/REF 82164; 10) 85124 CH12 40cm Onli Coude Hydrophilic Urethral Intermittent Catheter, SKU/REF 85124; 11) 85144 CH14 40cm Onli Coude Hydrophilic Urethral Intermittent Catheter, SKU/REF 85144; 12) 85164 CH16 40cm Onli Coude Hydrophilic Urethral Intermittent Catheter, SKU/REF 85164.

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

SKU/REF 82081, UDI/DI 00610075072354, Lot Numbers: 5F273, 5F283, 5H293, 5J153, 6C313. SKU/REF 82101, UDI/DI 00610075072361, Lot Numbers: 5F283, 5G013, 5K253, 6A303. SKU/REF 82121, UDI/DI 00610075072385, Lot Numbers: 5H283, 5H293, 5K183, 5K193, 5K213, 6A253, 6A263, 6B053. SKU/REF 82141, UDI/DI 00610075072392, Lot Numbers: 5E063, 5E073, 5E083, 5F193, 5F203, 5F213, 5F223, 6B133. SKU/REF 82084, UDI/DI 00610075072408, Lot Numbers: 5E293, 5G213, 5G223, 6A223, 6A213. SKU/REF 82104, UDI/DI 00610075072415, Lot Numbers: 5E293, 5E303, 5G243, 5G253, 5J063, 5J073, 5L153, 6D203, 6E293. SKU/REF 82124, UDI/DI 00610075072422, Lot Numbers: 5E303, 5F033, 5F063, 5F043, 5F053, 5J223, 5J283, 5J243, 5K143, 6A163, 6A153, 6B173, 6B183, 5F073. SKU/REF 82144, UDI/DI 00610075072439, Lot Numbers: 6A163, 5F113, 5F133, 5I073, 5G083, 5E243, 5E253, 5E233, 5F103, 5I123, 5F143, 5E263, 5E273, 5I113, 5E283, 5F123, 5F153, 5G093, 5G183, 5G173, 5I153, 5I133, 5I143, 5I063, 6A173, 6B243, 5I103, 6B223, 6B233, 6B253, 6B263, 6B213, 6C033, 6C043, 6B273, 6C013, 6C023, 6C053, 6B283. SKU/REF 82164, UDI/DI 00610075072446, Lot Numbers: 6A223, 6B183, 5F153, 5F163, 5F183, 5F173, 5G103, 5G143, 5G113, 5G153, 6A233, 6C253, 6C243, 6C223, 6C233, 6D303, 6D293, 6D283. SKU/REF 85124, UDI/DI 00610075097432, Lot Numbers: 5K143, 5K133, 6B193. SKU/REF 85144, UDI/DI 00610075097449, Lot Numbers: 6B263, 6B273, 6C253, 5K133, 5K123, 5K113, 5K053, 5K043, 5K063, 5K073, 6A113, 6A133, 5K103, 6A123, 6A093, 6A103, 6C283, 6C263, 6C273, 6E013. SKU/REF 85164, UDI/DI 00610075097456, Lot Numbers: 6D303, 5K133, 6B193, 6A093, 6C283, 5L053, 5L043, 6A083, 5L193, 6A023, 5L223, 6A053, 6C293.

Devices named in this recall

12 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines