LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor… recallZ-3081-2026 · Boston Scientific Corporation · initiated Aug 6, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Boston Scientific Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is potential for a delay in AF Burden Alert notification affecting the server software which supports the Insertable Cardiac Monitors (ICMs). Delayed notification of an AF Burden Alert may delay clinical evaluation or therapeutic intervention of an atrial arrhythmia.
FDA root cause category: Software design
What the company is doing
An Urgent Medical Device Correction dated August 2026 was sent to customers. Boston Scientific is notifying you of the potential for a delay in AF Burden Alert notification affecting the LATITUDE" Clarity server software which supports the LUX-Dx" family of ICMs. This delayed AF Burden Alert notification behavior by LATITUDE Clarity may occur if the Alert for AF Burden e Threshold is configured to ">0 hours per day" setting (a non-default configuration). The ICM is performing as intended with continued AF detection, data storage, and transmission of device data. Boston Scientific has received one complaint related to this delayed notification behavior by the LATITUDE Clarity System. No patient injuries or deaths have been reported. A server software update is planned to correct this behavior. Configuration Options for Atrial Alert Notifications: Atrial arrhythmia alert notification preferences may vary, and AF configurations may be set for individual patients and/or group settings. Until updated server software is available, consider the following options: Change Threshold for AF Burden Alert Setting Enable AF Events Alert with Minimum Duration If Unchanged, Potential Delay in AF Burden Alert Notification " Boston Scientific does not recommend replacement of the implanted ICM. " Please share this notification with all HCPs within your organization who manage patients using the LUXDx ICM and LATITUDE Clarity System. Mitigation. The LATITUDE Clarity system is designed to detect atrial fibrillation through two independent mechanisms: AF Burden and AF Event Alerts. Until a software update to the LATITUDE Clarity server software is available, configuring AF Events or changing AF Burden Alerts settings can mitigate delays in full device interrogation and enable the LATITUDE Clarity system to display alert notification as intended. Additional Information. The Regulatory Authority of your country has been informed of this communication. Adverse events should be reported
Recall details
- Recall number
- Z-3081-2026
- Status
- Open, Classified
- Initiated by the company
- Aug 6, 2026
- Posted by the FDA
- Sep 2, 2026
- Recalling firm
- Boston Scientific Corporation
- Product code
- MXD
- FDA submissions
- K231328
- Distribution
- Worldwide - US Nationwide distribution.
Product description
LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor System, SERVER SW LAT CLARITY DS US, Model Number 7260, All Versions (Current 7.8)
Code information (models, lots, serial numbers)
UDI-DI: 00802526613838
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- LATITUDE® Clarity 7260Boston Scientific Corporation · Recorder, event, implantable cardiac, (with arrhythmia detection)1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.