Skip to content

Halyard TOTAL JOINT recallZ-3009-2026 · AVID Medical, Inc. · initiated Jul 9, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from AVID Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Labeling Deficiency: The medical kit labeling fails to identify that the statlock component contains natural rubber latex.

FDA root cause category: Labeling Change Control

What the company is doing

On July 9, 2026 Avid Medical ( subsidiary of Owen & Minor) issued a Urgent: Medical Device correction notification to affected consignees. Avid Medical ask consignees to take the following actions: 1. Identify, segregate, and quarantine all impacted product identified. 2. Add warning labels to affected AVID kits to ensure all users are aware of the actions to discard the impacted component prior to use. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. 3. Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. 4. Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. 5. Complete the attached Recall Response Form and return by July 24, 2026.

Recall details

Recall number
Z-3009-2026
Status
Open, Classified
Initiated by the company
Jul 9, 2026
Posted by the FDA
Aug 21, 2026
Recalling firm
AVID Medical, Inc.
Quantity
120 Kits
Product code
LRO
Distribution
US Nationwide distribution in the states of MS, TX.

Product description

Halyard TOTAL JOINT Model/Catalog Number: MSBM036-03

Code information (models, lots, serial numbers)

Lot Code: Lot Number: 1662335 UDI: 10809160457887

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 99527).

Sources and updates