PORTRAIT COMPUTE AND CORE SERVICES V1.2, Ref# 5700592 recallZ-3037-2026 · GE Medical Systems Information Technologies Inc · initiated Jul 20, 2026
Class II Open, Classified All recalls posted in August 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Medical Systems Information Technologies Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area.
FDA root cause category: Device Design
What the company is doing
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 7/20/2026 by letter delivered using traceable means. The notice explained the issue, potential risk, and requested the following: "Actions to be taken by Customer/User You can continue to use your Portrait Mobile Monitoring Solution with Portrait Core Services SW V1.0.5, V1.1 or V1.2. Please ensure appropriate alternative monitoring is established for patients before a software update is started. A User Manual addendum is included with this letter that includes specific instructions related to alternative monitoring. Please place a copy of the addendum in Appendix A with the User Manual for your Portrait Mobile Monitoring Solution with Portrait Core Services SW V1.0.5, V1.1 or V1.2 system in your facility. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records." For questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Recall details
- Recall number
- Z-3037-2026
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Jul 20, 2026
- Posted by the FDA
- Aug 27, 2026
- Recalling firm
- GE Medical Systems Information Technologies Inc
- Quantity
- 4 units
- Product code
- MWI
- FDA submissions
- K234130
- Distribution
- US: MI, NY and Australia, Czech Republic, Finland, Germany, Italy, Netherlands, Qatar, Saudi Arabia, Switzerland, United Arab Emirates, United Kingdom.
Product description
PORTRAIT COMPUTE AND CORE SERVICES V1.2, Ref# 5700592
Code information (models, lots, serial numbers)
GTIN 00198953056752, Serial Numbers: SWT25510001TA, SWT26110010TA, SWT26070001TA, SWT26120008TA.
Other devices under the same FDA submission
1 device cleared under K234130 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Portrait CSSXBGE Medical Systems Information Technologies, Inc. · TRANSMITTERS AND RECEIVERS, PHYSIOLOGICAL SIGNAL, RADIOFREQUENCY1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99536).
- Z-3036-2026 · Open, ClassifiedPORTRAIT CORE SERVICES SOFTWARE V1.1, Ref# 5700209
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page