Medline Convenience Kits containing BD ChloraPrep Clear -… recallZ-3071-2026 · Medline Industries, LP · initiated Jul 20, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Downstream recall - manufacturer recalled specific lots of BD ChloraPrep Clear - 1mL Applicators. Affected product may exhibit an open or incomplete seal on the packaging of the applicator. Use of potentially contaminated applicator may lead to localized skin or soft tissue infection, or systemic infection or infectious complications.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Firm began notifying consignees on 7/20/2026 via First-Class Mail and email. Medline Industries LP instructed customers to request stickers to over-label the affected convenience kits with a label indicating to remove the affected component and replace it with product from supply. No kits are to be returned.
Recall details
- Recall number
- Z-3071-2026
- Status
- Open, Classified
- Initiated by the company
- Jul 20, 2026
- Posted by the FDA
- Aug 31, 2026
- Recalling firm
- Medline Industries, LP
- Quantity
- 222,653 total
- Product code
- LRS
- Distribution
- US Nationwide distribution.
Product description
Medline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators IV START KIT LAB DRAW-SH CDS860014Q IV START KIT DYKM1000B IV START KIT LAB DRAW-SH CDS860014P ULTRASOUND GUIDED PIV START KI DYNDV2684
Code information (models, lots, serial numbers)
CDS860014Q UDI-DI 10198459629822 Lots 26ABR359 26BBP789 26CBL940 26DBK987 26FBK255 26FBK816 DYKM1000B UDI-DI 10889942508143 Lots 24KMD782 24LMH404 25AMD765 25BMD744 CDS860014P UDI-DI 10888277797819 Lots 24IBA155 24JBY045 24LBM014 25HBR509 25JBX951 DYNDV2684 UDI-DI 10198459185465 Lot 24LBP243
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. CDS860014PMedline Industries, Inc. · I.V. start kit3 recalls0 reports
- Medline Industries, Inc. DYKM1000BMedline Industries, Inc. · I.V. start kit3 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99561).
- Z-3067-2026 · Open, ClassifiedMedline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators PAIN PRE OP KIT DYKS1476A
- Z-3068-2026 · Open, ClassifiedMedline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators OPEN HEART CDS983348AA CDS983348AB
- Z-3069-2026 · Open, ClassifiedMedline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators BASIN PACK-LF DYNJ42828B EYE BASIN TRAY-LF DYNJ42832B
- Z-3070-2026 · Open, ClassifiedMedline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators BLOOD CULTURE PINK PERIPHERAL DYNDV1875 BOTTOM DRAWER ACC010972 DRAWER #4 ACC010479A DRAWER…
- Z-3072-2026 · Open, ClassifiedMedline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators BLOOD CULTURE KIT 2 W/CONNECT DYNDH1487A
- Z-3073-2026 · Open, ClassifiedMedline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators ADULT BLOOD CULTURE KIT DYNDH1444D BLOOD CULTR KT DIVRSN DEVICE DYNDH1866 BLOOD CULTURE KIT DYNDH…
- Z-3074-2026 · Open, ClassifiedMedline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators DELUXE PRE OP KIT DYKS2018 GENERAL PREOP KIT DYKS1584A PRE OP KIT DYKS1216I PRE-OP KIT DYKM13…
- Z-3076-2026 · Open, ClassifiedMedline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators LABOR CDS CDS981762B VAG DELIVERY ADMISSION BAG DYKA1054F
- Z-3077-2026 · Open, ClassifiedMedline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators C-SECTION PREP DYNJ32702B
- Z-3078-2026 · Open, ClassifiedMedline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators CATARACT PREOP KIT DYKS1583A
- Z-3079-2026 · Open, ClassifiedMedline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators ORTHO ASPIRATION KIT DYKM1636A
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page