Medline Convenience Kits containing BD ChloraPrep Clear -… recallZ-3071-2026 · Medline Industries, LP · initiated Jul 20, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Downstream recall - manufacturer recalled specific lots of BD ChloraPrep Clear - 1mL Applicators. Affected product may exhibit an open or incomplete seal on the packaging of the applicator. Use of potentially contaminated applicator may lead to localized skin or soft tissue infection, or systemic infection or infectious complications.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Firm began notifying consignees on 7/20/2026 via First-Class Mail and email. Medline Industries LP instructed customers to request stickers to over-label the affected convenience kits with a label indicating to remove the affected component and replace it with product from supply. No kits are to be returned.

Recall details

Recall number
Z-3071-2026
Status
Open, Classified
Initiated by the company
Jul 20, 2026
Posted by the FDA
Aug 31, 2026
Recalling firm
Medline Industries, LP
Quantity
222,653 total
Product code
LRS
Distribution
US Nationwide distribution.

Product description

Medline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators IV START KIT LAB DRAW-SH CDS860014Q IV START KIT DYKM1000B IV START KIT LAB DRAW-SH CDS860014P ULTRASOUND GUIDED PIV START KI DYNDV2684

Code information (models, lots, serial numbers)

CDS860014Q UDI-DI 10198459629822 Lots 26ABR359 26BBP789 26CBL940 26DBK987 26FBK255 26FBK816 DYKM1000B UDI-DI 10889942508143 Lots 24KMD782 24LMH404 25AMD765 25BMD744 CDS860014P UDI-DI 10888277797819 Lots 24IBA155 24JBY045 24LBM014 25HBR509 25JBX951 DYNDV2684 UDI-DI 10198459185465 Lot 24LBP243

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 99561).

Sources and updates