Patient Information Center iX recallZ-3141-2026 · Philips North America · initiated Jul 31, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips North America or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to software malfunction, patient information system may freeze and become unresponsive.
FDA root cause category: Software design
What the company is doing
On July 31, 2026, Philips North America LLC issued an urgent medical device correction to affected consignees via USPS certified mail. Philips requested consignees to take the following actions: Users: 1. In cases where central monitoring has halted, please ensure that triaging of patients at bedside and a change in workflow occurs to more closely monitor patient status. Ensure audible alarm volume is on high in-patient rooms in case bedside devices disconnect and do not provide visual indicators that communication with the central. 2. Pass this notice to those who need to be aware within your organization or to any organization where affected product(s) have been potentially transferred. 3. Place this Urgent notification with the documentation of the Philips Patient Information Center (PIC) iX and associate devices in a place where it is most likely to be seen and viewed. 4. Implement the Technical Solution established by Philips as soon as available within the defined timeframe. Distributors: 1. Send the attached URGENT Medical Device Correction to each customer to whom you have distributed any affected product(s) 2. Perform a good faith effort to reach each customer a minimum of three (3) times, and if possible, using multiple contact methods. 3. Implement the Technical Solution established by Philips as soon as available within the timeframe communicated by Philips, and confirm with Philips the implementation of the correction of all affected product(s). In case any product(s) cannot be corrected, inform Philips of the reason why the Technical Solution could not be implemented.
Recall details
- Recall number
- Z-3141-2026
- Status
- Open, Classified
- Initiated by the company
- Jul 31, 2026
- Posted by the FDA
- Sep 11, 2026
- Recalling firm
- Philips North America
- Quantity
- 7359 Units
- Product code
- MHX
- FDA submissions
- K153702
- Distribution
- Worldwide - US Nationwide distribution including in the states and territories of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR and the countries of Andorra, Utd. Arab. Emir., Argentina, Austria, Australia, Aruba, Bosnia-Herz., Bangladesh, Belgium, Bulgaria, Bahrain, Bolivia, Brazil, Canada, Switzerland, Chile, Colombia, Costa Rica, Cyprus, Czech Republic, Germany, Denmark, Dominican Rep, Algeria, Estonia, Spain, Finland, Fiji, Faroe Islands, France, United Kingdom, Gibraltar, Guadeloupe, Greece, Hong Kong, Croatia, Hungary, Indonesia, Ireland, Israel, India, Iraq, Iceland, Italy, Jordan, Japan, Kenya, South Korea, Kuwait, Lebanon , Sri Lanka, Lithuania, Luxembourg, Latvia, Libya, Morocco, Monaco, Montenegro, Macedonia, Macao, Martinique, Maldives, Mexico, Malaysia, Namibia, Nicaragua, Netherlands, Norway, New Zealand, Oman, Panama, Peru, Philippines, Pakistan, Poland, Puerto Rico, Portugal, Qatar R¿union, Romania, Serbia, Russian Fed., Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, Thailand, Turkmenistan, Tunisia, T¿rkiye, Taiwan, Ukraine, Uzbekistan, Vietnam, South Africa.
Product description
Philips Patient Information Center iX REF:866389
Code information (models, lots, serial numbers)
All PIC iX software versions 4.0-4.5/UDI: (01)00884838127067
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Patient Information Center iX 4.4Philips Medizin Systeme Böblingen GmbH · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)1 recall7 deaths84 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 99566).
- Z-3142-2026 · Open, ClassifiedPhilips PIC iX Essentials REF:867093
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.