Skip to content

HOPKINS recallZ-3333-2026 · Karl Storz Endoscopy · initiated Aug 12, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Karl Storz Endoscopy or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to incorrect inside packaging label for eyepiece.

FDA root cause category: Employee error

What the company is doing

On 08/12/2026, an "URGENT: MEDICAL DEVICE RECALL" Letter was sent via email informing customers that, KARL STORZ distributed 6 HOPKINS Telescope 30, 4mm REF: 10005BA between November 24, 2025, and December 15, 2025, that contains an optic (eyepiece) that is incorrectly etched with REF: 10005AA, which is a HOPKINS Telescope 0, 4 mm. The optic (eyepiece) inside the packaging is in fact a REF: 10005BA HOPKINS Telescope 30, 4mm, as the outer packaging indicates. Customers are instructed to: 1. Immediately check for impacted serial numbers in inventory and quarantine the incorrectly etched devices for return to Karl Storz. 2. Pass on this Urgent Medical Device Correction to all users of the products listed below (Table 1 and Table 2) and all other persons who need to be aware within your organization. 3. If customers have or may have distributed the products listed, identify and promptly notify those recipients with a copy of this notice and response form. 4. Return the completed Customer Reply Form by E-Mail to the indicated contact below within 15 calendar days from the date of receipt. 5. Report any incidents related to this issue to the manufacturer, dealer, or local representative and, if applicable, to the national competent authority, as this is important feedback. For questions - email to [email protected].

Recall details

Recall number
Z-3333-2026
Status
Open, Classified
Initiated by the company
Aug 12, 2026
Posted by the FDA
Oct 1, 2026
Recalling firm
Karl Storz Endoscopy
Quantity
6 eyepieces
Product code
EOB
FDA submissions
K945788
Distribution
U.S. Nationwide distribution in the states of AZ, CA and MO.

Product description

HOPKINS Telescope 30, 4mm REF 10005BA UDI-DI code: 04048551045121 KARL STORZ HOPKINS and HOPKINS II telescopes are to be used for visualization of the surgical site during diagnostic and therapeutic endoscopic procedures.

Code information (models, lots, serial numbers)

Material Number: 10005BA UDI-DI code: 04048551045121 Serial Numbers: 100011, 100012, 10000Y, and 10000Z

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page