NovaSeq 6000Dx including DRAGEN Server, Catalog Number:… recallZ-0068-2027 · Illumina, Inc. · initiated Sep 17, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Illumina, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Retrospectively reported correction for an instrument intended for targeted sequencing of DNA libraries from human genomic DNA used with in vitro diagnostic assays has a server with a security vulnerability.

FDA root cause category: Software design

What the company is doing

On 9/17/2024, and 1/16/2025 field safety notices were emailed to customers who were asked to take actions on webpage: https://support.illumina.com/support-content/bmc.html Customers were also asked to complete and return the Verification Form via email to [email protected] This email can also be used to ask questions or to report potentially compromised servers.

Recall details

Recall number
Z-0068-2027
Status
Open, Classified
Initiated by the company
Sep 17, 2024
Posted by the FDA
Oct 1, 2026
Recalling firm
Illumina, Inc.
Quantity
80
Product code
PFF
Distribution
Worldwide - US Nationwide distribution in the states of TX, KS, NC, CA, UT, KA, MA, NJ, MD and the countries of Spain, South Korea, Japan, Australia, Italy, United Arab Emirates, Sweden, United Kingdom, Czech Republic, Singapore, Austria, Poland, France, Germany, Finland, Switzerland, Slovakia.

Product description

NovaSeq 6000Dx including DRAGEN Server, Catalog Number: 20068232; Illumina DRAGEN Server for NextSeq 550Dx, Catalog Number: 20086130

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Catalog/UDI-DI/Serial Numbers: 20068232/00816270020637/ADX00020, ADX00114, ADX00121, ADX00124, ADX00147, ADX00151, ADX00139, ADX00105, ADX00103, ADX00108, ADX00104, ADX00106, ADX00101, ADX00115, ADX00111, ADX00112, ADX00117, ADX00119, ADX00116, ADX00118, ADX00125, ADX00126, ADX00131, ADX00120, ADX00128, ADX00138, ADX00142, ADX00143, ADX00144, ADX00150, ADX00146, ADX00141, ADX00133, ADX00134, ADX00122, ADX00127, ADX00130, ADX00135, ADX00140, ADX00136, ADX00137, ADX00148, ADX00149, ADX00152, ADX00153. 20086130/ACN7003517, ACN7003523, ACM5003172, ACMC007033, ACN7003530, ACM6002504, ACM5003179, ACM4005610, ACN7003516, ACN1003268, ACN1003269, ACN1003262, ACM6002503, ACN7003532, ACN1003255, ACN7003531, ACN7003518, ACN1003258, ACN1003248, ACM9002241, ACM5002946, ACM9002243, ACM6002498, ACM6002015, ACN1003254, ACN1003271, ACN1003257, ACM6002494, ACMC007031, ACMC007048, ACN1003259, ACN1003256, ACN8002480, ACO2000280, ACO7001263

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.