Brand Name: Tandem t:slim mobile application Product Name:… recallZ-3330-2026 · Tandem Diabetes Care, Inc. · initiated Jul 16, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Tandem Diabetes Care, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A beta (early-release) version of the iOS application, version 2.13 (TestFlight) had a software issue preventing successful pairing of the t:slim application with Abbott FreeStyle Libre 3 Plus (FSL3P) continuous glucose monitoring (CGM) sensors, resulting in the insulin pump being unable to establish a near-field communication (NFC) connection with the CGM and requiring FSL3P users to use their t:slim insulin pumps in manual mode. FSL3P users are unable to access their CGM's glucose values from within the t:slim app and cannot use their pumps in automated control (closed-loop) mode, which could lead to users experiencing hypoglycemia or hyperglycemia.
FDA root cause category: Software Design Change
What the company is doing
On 07/16/2026, an "URGENT MEDICAL DEVICE CORRECTION" communication was sent via certified mail to customers informing them that an unexpected behavior was identified in the t:slim iOS mobile application version 2.13 made available through TestFlight. The issue involves users unable to pair a compatible FreeStyle Libre 3 Plus sensor after updating to version 2.13. If this occurs, the application will incorrectly display an "Incompatible Sensor" message and prevent sensor use. Customer are advised to: 1. Discontinue use of iOS t:slim mobile application version 2.13 2. Uninstall version 2.13 from their mobile device. 3. Return to the current production version of the t:slim mobile application (version 2.12.1) available on the Apple App Store. 4. Scan the compatible FreeStyle Libre 3 Plus sensor with version 2.12.1 when starting their next sensor session. For Questions - contact Tandum Customer Support at 1-877-801-6901 or email [email protected] (24/7/365).
Recall details
- Recall number
- Z-3330-2026
- Status
- Open, Classified
- Initiated by the company
- Jul 16, 2026
- Posted by the FDA
- Sep 29, 2026
- Recalling firm
- Tandem Diabetes Care, Inc.
- Quantity
- 177
- Product code
- QFG
- FDA submissions
- K201214, K203234, K232380
- Distribution
- US Nationwide distribution.
Product description
Brand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application, iOS Model/Catalog Number: iOS software version 2.13 Software Version: iOS software version 2.13 UDI-DI: 00389152470740 Component: No
Code information (models, lots, serial numbers)
Lot Code: UDI-DI: 00389152470641 Software Version: 2.13 (iOS only)
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Tandem t:slim Mobile Application, Android 1014706Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump2 recalls0 reports
Other devices under the same FDA submission
1 device cleared under K201214, K203234, K232380 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Pharmacy Kit, Starter Pack, Tandem t:slim X2, Control IQ+ 1017064Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump8 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page