FilmArray Pneumonia Panel plus (Pneumoplus) recallZ-0149-2027 · BioFire Diagnostics, LLC · initiated Aug 12, 2026
Open, Classified All recalls posted in October 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from BioFire Diagnostics, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to potential contaiminated of spefic lots of pneumonia and joint infection panels that could result in false positive results
FDA root cause category: Under Investigation by firm
What the company is doing
On 09/03/2026, an "URGENT: MEDICAL DEVICE RECALL (CORRECTION)" Letter was emailed to customers informing them that, bioMrieux internally identified increased rates of KPC unexpected positives in Quality Control (QC) testing of the BIOFIRE FILMARRAY PN Panel, BIOFIRE FILMARRAY PNplus Panel, and BIOFIRE JI Panel. Internal testing reproduced KPC unexpected positive results, suggesting KPC nucleic acid contamination in reagent manufacturing at levels that can lead to false positive results. Customers are instructed to: " Examine inventory for the lots identified in this recall. " If a positive KPC result obtained when using the affected product is inconsistent with local resistance epidemiology and clinical data, then the result should be confirmed with an alternative method. " If customer has further distributed or transferred this product to another facility, identify any recipients, notify them at once, and forward this notice. " Post the recall (correction) letter on or near affected products as a constant reminder of the actions that need to be taken. " Complete the accompanying Acknowledgement of Receipt Form and return to bioMrieux so that bioMrieux may acknowledge your receipt of this notification. For questions - contact your local Customer Service Representative at 1-800-682-2666, selecting option 1 for Technical Support and option 3 for BIOFIRE, or email [email protected]
Recall details
- Recall number
- Z-0149-2027
- Recall class
- Not yet classified by the FDA
- Status
- Open, Classified
- Initiated by the company
- Aug 12, 2026
- Posted by the FDA
- Oct 13, 2026
- Recalling firm
- BioFire Diagnostics, LLC
- Quantity
- 3,871
- Product code
- QDS
- FDA submissions
- K243222
- Distribution
- Worldwide - U.S. Nationwide distribution including in the states and territory of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The countries of United Arab Emirates, Netherland Antilles, Argentina, Austria, Bosnia and Herzegovina, Belgium, Bulgaria, Bolivia, Canada, Switzerland, Chile, China, Colombia, Costa Rica, Czech Republic, Germany, Denmark, Dominican Republic, Algeria, Ecuador, Spain, Finland, France, United Kingdom, Greece, Guatemala, Hong Kong, Honduras, Croatia, Hungary, Indonesia, Israel, India, Italy, Japan, Korea, Lebanon, Sri Lanka, Lithuania, Mexico, Malaysia, Nicaragua, Panama, Peru, Philippines, Poland, Portugal, Paraguay, Romania, Serbia, Singapore, El Salvador, Thailand, Taiwan, Venezuela, Vietnam.
Product description
Generic Name: FilmArray Pneumonia Panel plus Generic Name: FilmArray Pneumonia Panel plus Model/Catalog Number: RFIT-ASY-0143 (30 Kit), RFIT-ASY-0142 (6 Kit) Software Version: N/A Product Description: Multiple-type respiratory pathogen nucleic acid IVD, kit, nucleic acid technique (NAT) Component: N/A
Code information (models, lots, serial numbers)
Lot Code: UDI Nos 00815381020314 (30 Kit), 00815381020321 (6 Kit); Model Nos RFIT-ASY-0143 (30 Kit), RFIT-ASY-0142 (6 Kit); Lot Numbers 0638426, 0638526, 0676426, 0676526, 0679726, 0804826, 0804926, 0805026, 0883126, 0922126, 0941426, 0941526, 0997926, 0998026, 1020626, 1020726, 1077326, 1105326, 1118926, 1119026, 1119526, 1219226, 1239726, 1286526, 1286626, 1287326, 1287426, 1346226, 1346326, 1346426, 1403026, 1403126, 1403226, 1419326, 1419426, 1437626, 1479026, 1560626, 1560726, 1611826, 1611926, 1656926, 1683826, 1859726
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- FilmArray® Pneumonia Panel plus (Pneumoplus) RFIT-ASY-0143Biofire Diagnostics, LLC · Mers-Cov And Common Respiratory Pathogens Semi-Quantitative And Quantitative Multiplex Nucleic Acid Detection System4 recalls3 deaths4 reports
- FilmArray® Pneumonia Panel plus (Pneumoplus) RFIT-ASY-0142Biofire Diagnostics, LLC · Mers-Cov And Common Respiratory Pathogens Semi-Quantitative And Quantitative Multiplex Nucleic Acid Detection System1 recall0 reports
Product lines
Other devices under the same FDA submission
2 devices cleared under K243222 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- FilmArray® Pneumonia Panel (Pneumo) RFIT-ASY-0144Biofire Diagnostics, LLC · Respiratory Panel6 recalls7 reports
- FilmArray® Pneumonia Panel (Pneumo) RFIT-ASY-0145Biofire Diagnostics, LLC · Respiratory Panel3 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99592).
- Z-0148-2027 · Open, ClassifiedGeneric Name: FilmArray Pneumonia Panel Generic Name: FilmArray Pneumonia Panel Model/Catalog Number: RFIT-ASY-0144 (30 pack), RFIT-ASY-0145 (6 pack), FLM1-ASY-0150 (Verification…
- Z-0150-2027 · Open, ClassifiedGeneric Name: BioFire Joint Infection (JI) Panel Generic Name: BioFire Joint Infection (JI) Panel Model/Catalog Number: RFIT-ASY-0138 (30 Kit), RFIT-ASY-0135 (6 Kit) Software Versi…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page