Skip to content

Brand Name: Heidelberg Engineering GmbH recallZ-3215-2026 · Heidelberg Engineering GmbH · initiated Aug 6, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Heidelberg Engineering GmbH or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Wheels may come loose when module is being pushed or pulled during movement or transportation, affecting stability of the stand. The stand may tip over as a result. Falling stand may cause impact or crushing injuries, particularly to the feet, legs, hands or arms of the operator or nearby persons.

FDA root cause category: Component design/selection

What the company is doing

Firm began notifying US customers on August 6, 2026 via email. Instruction to Customers: There is a possibility that the castor wheels of the Flex stand are not properly attached. The wheels may come loose when the Flex Module is being pulled or pushed during movement or transportation, affecting the stability of the stand. As a result, the Flex stand may tip over if not correctly handled or supported by the operator. Users shall schedule a service appointment with your Heidelberg Engineering service partner as soon as possible so that the castor fastenings can be inspected and, if necessary, corrected.

Recall details

Recall number
Z-3215-2026
Status
Open, Classified
Initiated by the company
Aug 6, 2026
Posted by the FDA
Sep 18, 2026
Recalling firm
Heidelberg Engineering GmbH
Quantity
87
Product code
OBO
FDA submissions
K181594
Distribution
Worldwide - US Nationwide distribution in the state of Massachusetts and the countries of Australia, Costa Rica, Germany, France, India, Israel, Italy, Canada, Mexico, Netherlands, Austria, Philippines, Poland, Saudi Arabia, Switzerland, Spain. South Africa. Thailand, Czech Republic, United Kingdom, Vietnam.

Product description

Brand Name: Heidelberg Engineering GmbH Product Name: SPECTRALIS with Flex Module Model/Catalog Number: S3610, S2610, S2000, S1610 multimodal imaging platform optimized for the posterior segment of the eye

Code information (models, lots, serial numbers)

UDI-DI 04059237000235 Lot Code: Spec-07265, Spec-11121, Spec-11281, Spec-12096, Spec-13305, Spec-14403, Spec-15282, Spec-15587, Spec-20071, Spec-20348, Spec-20351, Spec-20591, Spec-20605, Spec-23473, Spec-23941, Spec-23950, Spec-23975, Spec-24507, Spec-24558, Spec-24582, Spec-24611

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page