Kit: Epidural/Pain TQC 20/Cs, REF: 82465 Containing Sodium… recallZ-3106-2026 · Medical Action Industries, Inc. 306 · initiated Jul 23, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medical Action Industries, Inc. 306 or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Kits containing Sodium Chloride ampules subject to another recall are recalled due to potentially compromised sterility.
FDA root cause category: Nonconforming Material/Component
What the company is doing
On 7/23/2026, "Urgent: Medical Device Recall" notices were emailed to customers who were asked to do the following: 1) Identify, Segregate, and Quarantine all impacted kits. 2) Add warning labels to affected kits to ensure all users are aware of the actions to take prior to use. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. This label should be applied in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. If pre-printed labels are needed, please contact your sales representative. 3) REMOVE AND DISCARD the Sodium Chloride, 0.9%, USP 10 mL Flush ampules prior to use. 4) Customers (end-users): If you have received product from a distributor, please contact the distributor or your sales representative with any questions you may have. 5) Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. 6) Complete and return the Recall Response Form via email to [email protected] Report any quality problems or adverse events associated with the products listed in this notification to [email protected] If you have any questions, contact Regulatory Affairs Manager at email: [email protected].
Recall details
- Recall number
- Z-3106-2026
- Status
- Open, Classified
- Initiated by the company
- Jul 23, 2026
- Posted by the FDA
- Sep 9, 2026
- Recalling firm
- Medical Action Industries, Inc. 306
- Quantity
- 1,960
- Product code
- OGJ
- Distribution
- US distribution to states of: IA, VA, MO, KS, FL, OH and LS.
Product description
Kit: Epidural/Pain TQC 20/Cs, REF: 82465 Containing Sodium Chloride Injection, 0.9%, USP 10ml ampules
Code information (models, lots, serial numbers)
UDI-DI: 10809160367414. Lot(Expiration Date): 322200(4/1/2026), 324586(4/1/2026), 327335(4/1/2026), 329618(9/1/2026), 347199(11/1/2026), 337283(11/1/2026), 350354(11/30/2026), 375449(5/31/2027)
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Medical Action Industries 82465Medical Action Industries Inc. · Nerve block tray1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99613).
- Z-3107-2026 · Open, ClassifiedKit: Single Shot Epidural 20/Cs, REF: 59553; Kit: Single Shot Epidural 20/Cs, REF: 83945B. Containing Sodium Chloride Injection, 0.9%, USP 10ml ampules
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page