Inca recallZ-3139-2026 · CooperSurgical, Inc. · initiated Aug 11, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from CooperSurgical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and lead to decreased oxygenation until connection is restored.
FDA root cause category: Device Design
What the company is doing
CooperSurgical notified consignees on about 08/11/2026 via "URGENT Medical Device Recall" letter. Consignees were notified of the issue and potential hazard, and were instructed to identify affected units on hand, if any loose connection is recognized during the use of the CPAP, replace the tubing with another tubing or CPAP device, and remove all product of the subject lots from inventory and note the number of units to be returned on the attached Recall Response Form that is provided. Distributors were instructed to remove all product of the subject lots from inventory and note the number of units to be returned on the attached Distributor Only Acknowledgement Form. Additionally, they were instructed to communicate the recall notification to all customers that they distributed these devices to.
Recall details
- Recall number
- Z-3139-2026
- Status
- Open, Classified
- Initiated by the company
- Aug 11, 2026
- Posted by the FDA
- Sep 18, 2026
- Recalling firm
- CooperSurgical, Inc.
- Quantity
- 2,643 boxes (13,215 units)
- Product code
- BZD
- FDA submissions
- K911780
- Distribution
- Nationwide distribution to CT, FL, GA, IA, IL, IN, MD, MI, MS, NJ, NY, OH, PA, VA. International distribution to Brazil and Peru.
Product description
INCA (Infant Nasal Cannulae Assy) Infant Nasal CPAP, Complete Set. Stay-Flex tubing component. 5 units per box. Model Numbers: 44-2707, 44-2709, 44-2710, 44-2712, 44-2715.
Code information (models, lots, serial numbers)
1. Model Number: 44-2707. UDI-DI: (01)20888937009433. Lot Code: 550019722. 2. Model Number: 44-2709. UDI-DI: (01)60888937009448. Lot Codes: 550031617, 550033781, 550031621. 3. Model Number: 44-2710. UDI-DI: (01)20888937009457. Lot Codes: 550031615, 550031616, 550031620, 550034760. 4. Model Number: 44-2712. UDI-DI: (01)20888937009464. Lot Codes: 550031611, 550031614, 550033779, 550031619. 5. Model Number: 44-2715. UDI-DI: (01)20888937009471. Lot Codes: 550031618, 550033780, 550034763.
Devices named in this recall
5 devices in the FDA device registry whose identifier appears in the recall's code information.
- Inca 44-2707Coopersurgical, Inc. · Ventilator, Non-Continuous (Respirator)1 recall0 reports
- Inca 44-2709Coopersurgical, Inc. · Ventilator, Non-Continuous (Respirator)1 recall0 reports
- Inca 44-2710Coopersurgical, Inc. · Ventilator, Non-Continuous (Respirator)1 recall0 reports
- Inca 44-2712Coopersurgical, Inc. · Ventilator, Non-Continuous (Respirator)1 recall0 reports
- Inca 44-2715Coopersurgical, Inc. · Ventilator, Non-Continuous (Respirator)1 recall0 reports
Product lines
Other devices under the same FDA submission
4 devices cleared under K911780 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Inca 44-0707Coopersurgical, Inc. · Ventilator, Non-Continuous (Respirator)1 recall0 reports
- Inca 44-0709Coopersurgical, Inc. · Ventilator, Non-Continuous (Respirator)1 recall0 reports
- Inca 44-0712Coopersurgical, Inc. · Ventilator, Non-Continuous (Respirator)1 recall0 reports
- Inca 44-0715Coopersurgical, Inc. · Ventilator, Non-Continuous (Respirator)1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99643).
- Z-3140-2026 · Open, ClassifiedINCA (Infant Nasal Cannulae Assy) Neonatal Nasal CPAP, Replacement Set. Stay-Flex tubing component. 5 units per box. Model Numbers: 44-0712, 44-0707, 44-0709, 44-0715.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.