Skip to content

CHARLOTTE CLAW Screw, REF: 40122716, intended for use with… recallZ-3334-2026 · Wright Medical Technology, Inc. · initiated Aug 18, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Wright Medical Technology, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Screws, intended for use with polyaxial compression plating system intended for foot osteotomies, was manufactured with missing threads on the screwheads, which if used would prevent a secure rigid engagement with the plate's internal threads, which may lead to surgery prolongation to retrieve another device, or alternative surgery method if no replacement device is readily available.

FDA root cause category: Process control

What the company is doing

On 8/13/2026, recall notices were mailed and emailed to customers who were asked to do the following: 1) Review the product issue and risks as communicated in this Recall Notice and communicate the issue as appropriate in your organization. 2) Sign and return the enclosed business reply form by email to [email protected] 3) Upon receipt of the completed business reply form, firm will arrange for product return. 4) If you have further distributed the affected product, please notify the applicable parties at once about this recall. You may copy and distribute this notification letter. a. Inform firm if any of the subject devices have been distributed to other organizations, including contact details, so that firm can inform the recipients appropriately. b. If you are a distributor, notify your affected customers. For questions or concerns, contact the firm at [email protected]

Recall details

Recall number
Z-3334-2026
Status
Open, Classified
Initiated by the company
Aug 18, 2026
Posted by the FDA
Oct 1, 2026
Recalling firm
Wright Medical Technology, Inc.
Quantity
34
Product code
HRS
FDA submissions
K133715
Distribution
US Nationwide distribution in the states of FL, TX, PA, CA, ID, MI, OK, TN, NJ, WI, CO.

Product description

CHARLOTTE CLAW Screw, REF: 40122716, intended for use with CLAW II Polyaxial Compression Plating System

Code information (models, lots, serial numbers)

UDI-DI: 00840420109705, Lot: 1826666, Distributed: 18-Aug-2025 through 09-Apr-2026

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates