Pi-Cardia Inc recallZ-3213-2026 · Pi Cardia Ltd · initiated Jul 6, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Pi Cardia Ltd or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Transfemoral catheter, positing arm may detached due to weld issue.
FDA root cause category: Process control
What the company is doing
On September 3, 2026 Pi-Cardia issued a Urgent Medical Device Recall Update notification to affected consignees via email. This update provided additional information on the risk to health as well as consignee instructions. Original notification was issued July 9, 2026, Pi Cardia Ltd. informed consignees the following: 1. A member of the Pi-Cardia team will contact you promptly with instructions regarding identification and return of any affected inventory. 2. If you have questions in the meantime, please contact your Pi-Cardia representative. 3. Pi-Cardia will provide additional updates as appropriate.
Recall details
- Recall number
- Z-3213-2026
- Status
- Open, Classified
- Initiated by the company
- Jul 6, 2026
- Posted by the FDA
- Sep 17, 2026
- Recalling firm
- Pi Cardia Ltd
- Quantity
- 46 units
- Product code
- SCZ
- Distribution
- US Nationwide distribution in the states of Utah, Kansas, California, Massachusetts, Pennsylvania, New Hampshire, Delaware, and New York.
Product description
Pi-Cardia Inc. ShortCut REF: Z-A-20.003-US
Code information (models, lots, serial numbers)
UDI-DI - (01)7290020101242 All serial numbers in the affected lot SCUS-00013: SC919, SC920, SC921, SC923, SC924, SC925, SC926, SC927, SC928, SC929, SC930, SC931, SC932, SC933, SC934, SC935, SC936, SC937, SC938, SC939, SC940, SC941, SC942, SC943, SC944, SC945, SC946, SC947, SC948, SC949, SC950, SC951, SC952, SC953, SC954, SC956, SC957, SC958, SC960, SC961, SC962, SC963, SC964, SC966, SC967 and SC968.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- ShortCut Z-A-20.003-USPi-Cardia Ltd · Percutaneous Catheter For Cutting Or Splitting Heart Valve Leaflets Concomitant To Transcatheter Valve Procedures1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.