Skip to content

Convenience kits recallZ-3209-2026 · Medline Industries, LP · initiated Jul 31, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Kits may exhibit open seals in sterile pouch packaging. Open seal may compromise product sterility. If not detected prior to use, issue may lead to contamination of sterile field or patient infection if product is used.

FDA root cause category: Under Investigation by firm

What the company is doing

Firm began notifying customers on July 31, 2026 via email and first-class mail. Customers were instructed to discard any affected sterile convenience kits.

Recall details

Recall number
Z-3209-2026
Status
Open, Classified
Initiated by the company
Jul 31, 2026
Posted by the FDA
Sep 16, 2026
Recalling firm
Medline Industries, LP
Quantity
60
Product code
LRO
FDA submissions
K213481
Distribution
US Nationwide distribution in states of Arkansas, Florida, North Carolina, Maryland, and Texas.

Product description

Convenience kits MYRINGOTOMY PACK-LF DYNJ59097A

Code information (models, lots, serial numbers)

UDI-DI 10198459046551 Lot 25FBU754

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

321 devices cleared under K213481 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 99667).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page