Pipeline Flex Embolization Device with Shield Technology… recallZ-3326-2026 · Micro Therapeutics, Inc. · initiated Jun 25, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Micro Therapeutics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to incorrect label/product dimensions. Product is supposed to be 3.00 mm in diameter, but unit is packaging was a 2.50 mm in diameter.
FDA root cause category: Under Investigation by firm
What the company is doing
On 06/25/2026, an "URGENT: MEDICAL DEVICE RECALL" letter was sent to customers to inform them that, Medtronic has confirmed a labeling error involving implant diameter in the affected lot. Devices from Lot # D112860 labeled as PED2-300-12 (3.00 mm diameter, 12 mm length) contain 2.50 mm diameter implants instead of the labeled 3.00 mm diameter. Customers are instructed to: 1. Do NOT use any impacted Pipeline Flex Embolization Device with Shield Technology devices listed. Remove and quarantine all unused impacted product units in their inventory. 2. Return the impacted product(s) to Medtronic. Their Medtronic representative can assist in facilitating the return of product and/or identifying suitable replacement product as necessary. 3. Complete the attached Customer Confirmation Form and email it to [email protected]. 4. Review their records to identify any patients who may have been treated with devices from the affected lot and assess whether any follow-up is appropriate. 5. Share this recall communication within their organization, with other organizations where impacted devices have been transferred, and any other associated organizations that may be impacted by this action. Maintain a copy of this letter for their records and the records of their patients with Pipeline Shield. For questions - contact Medtronic at 1 (800) 633-8766 or (763) 514-4000 (Worldwide) or Email: [email protected]
Recall details
- Recall number
- Z-3326-2026
- Status
- Open, Classified
- Initiated by the company
- Jun 25, 2026
- Posted by the FDA
- Sep 28, 2026
- Recalling firm
- Micro Therapeutics, Inc.
- Quantity
- 2
- Product code
- OUT
- FDA submissions
- P100018S026
- Distribution
- U.S. Nationwide distribution in the states of GA and TX.
Product description
Pipeline Flex Embolization Device with Shield Technology REF PED2-300-12 UDI-DI code: 00763000448806 he Pipeline Flex Embolization Device with Shield Technology consists of a permanent implant combined with a guidewire based delivery system. The Pipeline Flex Embolization Device with Shield Technology implant is a braided, multi-alloy, mesh cylinder woven from platinum/tungsten and cobalt-chromium-nickel alloy wires and indicated for the endovascular treatment of adults (22 years of age or older) with large or giant wide-necked intracranial aneurysms (IAs) in the internal carotid artery from the petrous to the superior hypophyseal segments.
Code information (models, lots, serial numbers)
Product/Catalog Number: PED2-300-12 UDI-DI/GTIN: 00763000448806 Lot Number: #D112860
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Pipeline TM Flex Embolization Device with Shield Technology TM PED2-300-12Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter2 recalls1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page