Trilogy Delivery Catheter recallZ-0135-2027 · Jenavalve Technology Inc. · initiated Jul 31, 2026
Open, Classified All recalls posted in October 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Jenavalve Technology Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Delivery Catheter may have sleeve liner separation that results in difficulty or inability to deploy heart valve device.
FDA root cause category: Nonconforming Material/Component
What the company is doing
On the dates of August 1, 4, 6, and 19th, JenaValve Technology Inc. issued a Urgent: Medical Device Recall notification via email. JenaValve asked consignees to take the following actions: 1. Review the information contained in this letter and ensure that anyone in your facility who needs to be aware of this notification reads this letter carefully. 2. Examine your inventory immediately to determine if you have Trilogy Delivery Catheters from the affected lots and quarantine the subject products. 3. JenaValve field team members will schedule a visit to retrieve the subject Trilogy Delivery Catheters. If you prefer to return the products yourself, please contact us at 866-588-8287 and we will provide you with an RGA number prior to you shipping them to: JenaValve Technology Inc., 4 Cromwell, Irvine, California 92618, USA; Attn: Quality Department; Reference: FA26001 4. Please complete and sign the Business reply form (BRF) of this letter and email it to [email protected] or give it to your JenaValve field representative. 5. Upon receipt of the completed BRF and returned product, JenaValve will provide replacement products. Incomplete BRFs cannot be processed. 6. If any of the subject Trilogy Delivery Catheters have been transferred to another facility, identify the receiving facility on the Business Reply Form, contact [email protected], and provide a copy of this letter to the relevant personnel.
Recall details
- Recall number
- Z-0135-2027
- Recall class
- Not yet classified by the FDA
- Status
- Open, Classified
- Initiated by the company
- Jul 31, 2026
- Posted by the FDA
- Oct 9, 2026
- Recalling firm
- Jenavalve Technology Inc.
- Quantity
- 108 units
- Product code
- NPT
- FDA submissions
- P250024
- Distribution
- US Nationwide distribution in the states of AZ, CA, FL, GA, IL, MA, MD, MI, MN, MO, NY, OH, PA, SC, TX, UT, VA, WA and the countries of Germany, Denmark, Great Britain, Italy.
Product description
JenaValve Trilogy Delivery Catheter, Size 25MM REF: TRILOGY-DC-M
Code information (models, lots, serial numbers)
GTIN - 00850000898141/Lot # DS-040826-005, DS-040926-002, DS-042326-001, DS-042826-007, DS-042926-008, DS-050826-005, DS-051526-001
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Trilogy Delivery Catheter TRILOGY-DC-MJenavalve Technology, Inc. · Aortic Valve, Prosthesis, Percutaneously Delivered1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 99697).
- Z-0134-2027 · Open, ClassifiedJenaValve Trilogy Delivery Catheter, Size 23MM REF: TRILOGY-DC-S
- Z-0136-2027 · Open, ClassifiedJenaValve Trilogy Delivery Catheter, Size 27MM REF: TRILOGY-DC-L
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page