Skip to content

Presource Convenience Kits recallZ-3254-2026 · Cardinal Health 200, LLC · initiated Aug 10, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Cardinal Health 200, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The kits contain applicators that may exhibit an open or incomplete seal on the packaging of the applicator. Use of a potentially contaminated applicator could result in no patient harm or may lead to localized skin or soft tissue infection. Less commonly, systemic infection or other infectious complications may occur.

FDA root cause category: Nonconforming Material/Component

What the company is doing

An URGENT: PRODUCT CORRECTION notification letter dated 8/10/26 was sent to customers. If you locate affected product within your facility, please do the following: 1. QUARANTINE affected kits. 2. ALERT your clinicians of this action. 3. AFFIX a WARNING LABEL to the front of each Presource kit so that it is clearly visible to clinicians, instructing them to remove and discard the recalled component(s). 4. RETURN a copy of the enclosed Acknowledgment Form confirming your receipt of this action and the completion of labeling product via fax to 614.652.9648. Additional labels will be sent to your facility upon request. Please send an email to: [email protected] For component credit or replacement, please contact the following: " Hospitals Cardinal Customer Service at 866.551.0533 " Distributors Distributor Management at 800.635.6021 " Federal Government facilities Government Customer Service at 800.444.1166 If you are a Distributor and have further distributed affected kits please notify your customers of this recall. Should you have any questions, require additional labels or desire special assistance relating to this recall, please feel free to contact Cardinal Health Quality Systems at 800.292.9332.

Recall details

Recall number
Z-3254-2026
Status
Open, Classified
Initiated by the company
Aug 10, 2026
Posted by the FDA
Sep 22, 2026
Recalling firm
Cardinal Health 200, LLC
Quantity
17503 kits
Product code
LRP
Distribution
US Nationwide distribution in the states of FL, KY, MO, NC, OH, TX.

Product description

Presource Convenience Kits: 1) KIT, PRE OP , Catalog Number: AMN548616A; 2) KIT, NICU ADMIT NEW HAN , Catalog Number: BA1DH0116B; 3) KIT, PRE PROCEDURE , Catalog Number: IVSJV8302E; 4) KIT, IV SUPPLIES , Catalog Number: IVS3223531; 5) KIT, IV SUPPLIES , Catalog Number: IVS3223532; 6) KIT, IV SUPPLIES , Catalog Number: IVS3223533; 7) KIT, IV SUPPLIES , Catalog Number: IVS3223534; 8) KIT, IV SUPPLIES , Catalog Number: IVS3223535; 9) KIT, IV SUPPLIES , Catalog Number: IVS322358M; 10) KIT, IV SUPPLIES , Catalog Number: IVS322358N; 11) KIT, IV SUPPLIES , Catalog Number: IVS322358O; 12) KIT, PATIENT ADMISSIONS , Catalog Number: AMN333032A; 13) KIT, C SECTION , Catalog Number: MAT42267AL; 14) KIT, C SECTION , Catalog Number: MAT4226725; 15) KIT, VAGINAL DELIVERY , Catalog Number: MAT422677G; 16) KIT, VAGINAL DELIVERY , Catalog Number: MAT4226792; 17) KIT, VAGINAL DELIVERY , Catalog Number: MAT4226793; 18) KIT, VAGINAL DELIVERY , Catalog Number: MAT4226794; 19) KIT, VAGINAL DELIVERY , Catalog Number: MAT4226795

Code information (models, lots, serial numbers)

Show the full code information (10k characters)

1) AMN548616A, UDI-DI: 10197106721834(each), 10197106721834(case), Lot Number: 4249818 ; 2) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 5124984 ; 3) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 5129488 ; 4) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 5129488 ; 5) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 5188103 ; 6) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 5188103 ; 7) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 5211212 ; 8) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 5155824 ; 9) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 5211985 ; 10) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 5279226 ; 11) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 5155824 ; 12) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 5188103 ; 13) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 5211212 ; 14) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 5246952 ; 15) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 5279226 ; 16) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 5279226 ; 17) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 6062168 ; 18) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 5155824 ; 19) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 5279226 ; 20) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 6062168 ; 21) BA1DH0116B, UDI-DI: 0(each), 0(case), Lot Number: 6051588 ; 22) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5124984 ; 23) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5124984 ; 24) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5129488 ; 25) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5124984 ; 26) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5129488 ; 27) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5129488 ; 28) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5188103 ; 29) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5195818 ; 30) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5211907 ; 31) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5211985 ; 32) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5211212 ; 33) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5211985 ; 34) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5279226 ; 35) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5129488 ; 36) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5211985 ; 37) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 6037918 ; 38) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5129488 ; 39) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5155824 ; 40) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5279226 ; 41) IVS3223531, UDI-DI: 10198956173941(each), 10198956173941(case), Lot Number: 5124984 ; 42) IVS3223532, UDI-DI: 10198956173958(each), 10198956173958(case), Lot Number: 5124984 ; 43) IVS3223533, UDI-DI: 10198956362536(each), 10198956362536(case), Lot Number: 5188103 ; 44) IVS3223533, UDI-DI: 10198956362536(each), 10198956362536(case), Lot Number: 5211985 ; 45) IVS3223534, UDI-DI: 10198956470347(each), 10198956470347(case), Lot Number: 5129488 ; 46) IVS3223534, UDI-DI: 10198956470347(each), 10198956470347(case), Lot Number: 5155824 ; 47) IVS3223534, UDI-DI: 10198956470347(each), 10198956470347(case), Lot Number: 5310701 ; 48) IVS3223535, UDI-DI: 10198956527010(each), 10198956527010(case), Lot Number: 5188103 ; 49) IVS3223535, UDI-DI: 10198956527010(each), 10198956527010(case), Lot Number: 60622168 ; 50) IVS322358M, UDI-DI: 10195594809362(each), 10195594809362(case), Lot Number: 4249818 ; 51) IVS322358M, UDI-DI: 10195594809362(each), 10195594809362(case), Lot Number: 4248961 ; 52) IVS322358M, UDI-DI: 10195594809362(each), 10195594809362(case), Lot Number: 4249818 ; 53) IVS322358M, UDI-DI: 10195594809362(each), 10195594809362(case), Lot Number: 4249818 ; 54) IVS322358M, UDI-DI: 10195594809362(each), 10195594809362(case), Lot Number: 4249818 ; 55) IVS322358N, UDI-DI: 10198956106086(each), 10198956106086(case), Lot Number: 5124984 ; 56) IVS322358N, UDI-DI: 10198956106086(each), 10198956106086(case), Lot Number: 4249818 ; 57) IVS322358N, UDI-DI: 10198956106086(each), 10198956106086(case), Lot Number: 5124984 ; 58) IVS322358N, UDI-DI: 10198956106086(each), 10198956106086(case), Lot Number: 5129488 ; 59) IVS322358N, UDI-DI: 10198956106086(each), 10198956106086(case), Lot Number: 5188103 ; 60) IVS322358N, UDI-DI: 10198956106086(each), 10198956106086(case), Lot Number: 5188103 ; 61) IVS322358N, UDI-DI: 10198956106086(each), 10198956106086(case), Lot Number: 5195818 ; 62) IVS322358N, UDI-DI: 10198956106086(each), 10198956106086(case), Lot Number: 5188103 ; 63) IVS322358N, UDI-DI: 10198956106086(each), 10198956106086(case), Lot Number: 5188103 ; 64) IVS322358N, UDI-DI: 10198956106086(each), 10198956106086(case), Lot Number: 5129488 ; 65) IVS322358O, UDI-DI: 10198956517073(each), 10198956517073(case), Lot Number: 6051588 ; 66) AMN333032A, UDI-DI: 10198956367791(each), 10198956367791(case), Lot Number: 5248375 ; 67) AMN333032A, UDI-DI: 10198956367791(each), 10198956367791(case), Lot Number: 5309623 ; 68) AMN333032A, UDI-DI: 10198956367791(each), 10198956367791(case), Lot Number: 5309623 ; 69) AMN333032A, UDI-DI: 10198956367791(each), 10198956367791(case), Lot Number: 5310701 ; 70) AMN333032A, UDI-DI: 10198956367791(each), 10198956367791(case), Lot Number: 6051588 ; 71) AMN333032A, UDI-DI: 10198956367791(each), 10198956367791(case), Lot Number: 5150160 ; 72) AMN333032A, UDI-DI: 10198956367791(each), 10198956367791(case), Lot Number: 5188102 ; 73) AMN333032A, UDI-DI: 10198956367791(each), 10198956367791(case), Lot Number: 5188103 ; 74) AMN333032A, UDI-DI: 10198956367791(each), 10198956367791(case), Lot Number: 5350797 ; 75) MAT42267AL, UDI-DI: 10198956120129(each), 10198956120129(case), Lot Number: 5126398 ; 76) MAT42267AL, UDI-DI: 10198956120129(each), 10198956120129(case), Lot Number: 5126398 ; 77) MAT42267AL, UDI-DI: 10198956120129(each), 10198956120129(case), Lot Number: 5181074 ; 78) MAT42267AL, UDI-DI: 10198956120129(each), 10198956120129(case), Lot Number: 5188102 ; 79) MAT42267AL, UDI-DI: 10198956120129(each), 10198956120129(case), Lot Number: 5188102 ; 80) MAT42267AL, UDI-DI: 5276968 (each), 10198956120129(case), Lot Number: 10198956120129; 81) MAT4226725, UDI-DI: 5126398 (each), 10198956181557(case), Lot Number: 10198956181557; 82) MAT422677G, UDI-DI: 10198956120136(each), 10198956120136(case), Lot Number: 5126398 ; 83) MAT422677G, UDI-DI: 10198956120136(each), 10198956120136(case), Lot Number: 5126398 ; 84) MAT422677G, UDI-DI: 10198956120136(each), 10198956120136(case), Lot Number: 5129058 ; 85) MAT422677G, UDI-DI: 10198956120136(each), 10198956120136(case), Lot Number: 5126398 ; 86) MAT422677G, UDI-DI: 10198956120136(each), 10198956120136(case), Lot Number: 5210853 ; 87) MAT4226792, UDI-DI: 10198956207974(each), 10198956207974(case), Lot Number: 5129058 ; 88) MAT4226793, UDI-DI: 10198956426405(each), 10198956426405(case), Lot Number: 5350797 ; 89) MAT4226794, UDI-DI: 10198956484559(each), 10198956484559(case), Lot Number: 5309623 ; 90) MAT4226794, UDI-DI: 10198956484559(each), 10198956484559(case), Lot Number: 5310701 ; 91) MAT4226795, UDI-DI: 10198956538023(each), 10198956538023(case), Lot Number: 6051588

Devices named in this recall

17 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 99710).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page