BD Phoenix Panel NMIC-312. Catalog Number: 449082 recallZ-0139-2027 · Becton Dickinson & Co. · initiated Aug 13, 2026

Open, Classified All recalls posted in October 2026

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Becton Dickinson & Co. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for quality control failures with Klebsiella pneumoniae ATCC BAA-1705 when using the CPO Ambler classification feature.

FDA root cause category: Under Investigation by firm

What the company is doing

Beckton Dickenson (BD) notified consignees on about 08/13/2026 via mailed and emailed "URGENT: Medical Device Correction" letter. Clinical users were instructed to do the following: 1. Do not report CPO Ambler classification results generated from affected BD Phoenix CPO detect 6-well panels. 2. Do not release Ambler classification information from BD Phoenix panels on patient reports, laboratory information system reports, electronic health records, or other clinical reporting outputs. 3. CPO positive results, other resistance marker results, and any antimicrobial AST MIC and/or SIR result from the BD Phoenix panels are unaffected and may still be reported according to routine laboratory reporting practice. 4. Confirmatory testing to verify appropriate Ambler Classification should be determined and performed in accordance with established laboratory procedures. 5. There are no recommendations regarding results review or sample retesting. Customers were instructed to ensure the contents of the notification are read and understood, share and post the customer letter within each facility network, forward to any customers to whom product may have been distributed to. Additionally, customers were instructed to complete and return the provided Customer Response Form. BD is investigating the root cause and will implement appropriate corrective actions to rectify this issue as soon as possible. **UPDATE 09/02/2026: Firm expanded recall and sent updated notification to include an additional product code. Instructions remained the same.

Recall details

Recall number
Z-0139-2027
Recall class
Not yet classified by the FDA
Status
Open, Classified
Initiated by the company
Aug 13, 2026
Posted by the FDA
Oct 9, 2026
Recalling firm
Becton Dickinson & Co.
Quantity
553,625 units
Product code
LON
FDA submissions
K181665
Distribution
US Nationwide distribution.

Product description

BD Phoenix Panel NMIC-312. Catalog Number: 449082.

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

Catalog Number: 449082. UDI Numbers: (10)5324466(17)26 1116(30)0025(01)3 0382904490820, (10)5349874(17)26 1209(30)0025(01)3 0382904490820, (10)6008828(17)26 1226(30)0025(01)3 0382904490820, (10)5350977(17)26 1228(30)0025(01)3 0382904490820, (10)6028392(17)27 0120(30)0025(01)3 0382904490820, (10)6034513(17)27 0131(30)0025(01)3 0382904490820, (10)6042998(17)27 0202(30)0025(01)3 0382904490820, (10)6042021(17)27 0203(30)0025(01)3 0382904490820, (10)6056431(17)27 0217(30)0025(01)3 0382904490820, (10)6069033(17)27 0309(30)0025(01)3 0382904490820, (10)6083447(17)27 0327(30)0025(01)3 0382904490820, (10)6098295(17)27 0330(30)0025(01)3 0382904490820, (10)6104811(17)27 0412(30)0025(01)3 0382904490820, (10)6104042(17)27 0419(30)0025(01)3 0382904490820, (10)6118556(17)27 0421(30)0025(01)3 0382904490820, (10)6125526(17)27 0501(30)0025(01)3 0382904490820, (10)6128740(17)27 0503(30)0025(01)3 0382904490820, (10)6132650(17)27 0508(30)0025(01)3 0382904490820, (10)6132735(17)27 0510(30)0025(01)3 0382904490820, (10)6148198(17)27 0518(30)0025(01)3 0382904490820, (10)6148397(17)27 0524(30)0025(01)3 0382904490820, (10)6153088(17)27 0525(30)0025(01)3 0382904490820, (10)6153231(17)27 0526(30)0025(01)3 0382904490820, (10)6160397(17)27 0602(30)0025(01)3 0382904490820, (10)6160450(17)27 0605(30)0025(01)3 0382904490820, (10)6166259(17)27 0604(30)0025(01)3 0382904490820, (10)6166265(17)27 0606(30)0025(01)3 0382904490820, (10)6181978(17)27 0622(30)0025(01)3 0382904490820, (10)6181006(17)27 0623(30)0025(01)3 0382904490820, (10)6181015(17)27 0623(30)0025(01)3 0382904490820, (10)6181164(17)27 0626(30)0025(01)3 0382904490820, (10)6181151(17)27 0626(30)0025(01)3 0382904490820, (10)6181991(17)27 0622(30)0025(01)3 0382904490820, (10)6188614(17)27 0703(30)0025(01)3 0382904490820, (10)6188641(17)27 0703(30)0025(01)3 0382904490820, (10)6188683(17)27 0705(30)0025(01)3 0382904490820, (10)6188905(17)27 0710(30)0025(01)3 0382904490820, (10)6188922(17)27 0711(30)0025(01)3 0382904490820, (10)6195847(17)27 0713(30)0025(01)3 0382904490820, (10)6195877(17)27 0714(30)0025(01)3 0382904490820. Lot Numbers: 5324466, 5349874, 6008828, 5350977, 6028392, 6034513, 6042998, 6042021, 6056431, 6069033, 6083447, 6098295, 6104811, 6104042, 6118556, 6125526, 6128740, 6132650, 6132735, 6148198, 6148397, 6153088, 6153231, 6160397, 6160450, 6166259, 6166265, 6181978, 6181006, 6181015, 6181164, 6181151, 6181991, 6188614, 6188641, 6188683, 6188905, 6188922, 6195847, 6195877.

Other devices under the same FDA submission

2 devices cleared under K181665 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 99725).