Medline convenience kits containing Healthmark Instrument… recallZ-3229-2026 · Medline Industries, LP · initiated Aug 10, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Kits contain Healthmark Instrument Channel Brushes, which were recalled by the manufacturer due to the potential for a sharp edge/burr on the end of the brush. Potential hazards associated include cut injury/abrasion and infection.
FDA root cause category: Under Investigation by firm
What the company is doing
Firm began notifying consignees on August 10, 2026. Customers may continue to use Instrument Channel Brushes contained in Medline convenience kits in accordance with recommendations from the manufacturer of the brushes, Healthmark: Customers should be aware of possible sharp burrs when handling and utilizing Instrument Channel Brushes. If a sharp burr is discovered, discard or return the brush.
Recall details
- Recall number
- Z-3229-2026
- Status
- Open, Classified
- Initiated by the company
- Aug 10, 2026
- Posted by the FDA
- Sep 18, 2026
- Recalling firm
- Medline Industries, LP
- Quantity
- 90 kits (US)
- Product code
- LRO
- FDA submissions
- K213481
- Distribution
- US Nationwide distribution in the states of MN and SD.
Product description
Medline convenience kits containing Healthmark Instrument Channel Brushes. SINUS PACK DYNJ48258D and Sinus Pack DYNJ48258F.
Code information (models, lots, serial numbers)
DYNJ48258D UDI-DI 10889942563272 Lots 23CBH459 23GBJ128 24ABM400
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYNJ48258DMedline Industries, Inc. · Ear, nose, and throat surgical tray1 recall0 reports
Other devices under the same FDA submission
321 devices cleared under K213481 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Centurion ART690Centurion Medical Products Corporation · General Surgery Tray4 recalls0 reports
- Centurion CVI4505Centurion Medical Products Corporation · General Surgery Tray3 recalls0 reports
- MEDLINE ECVC6355Medline Industries, Inc. · General surgery tray3 recalls0 reports
- MEDLINE ECVC1045Medline Industries, Inc. · General surgery tray3 recalls0 reports
- Centurion ART255Centurion Medical Products Corporation · General surgery tray3 recalls0 reports
- Centurion CVI3880Centurion Medical Products Corporation · General surgery tray3 recalls0 reports
- MEDLINE ECVC1375Medline Industries, Inc. · General surgery tray3 recalls0 reports
- MEDLINE DYNDV2520Medline Industries, Inc. · General surgery tray3 recalls0 reports
- Centurion ST11670ACenturion Medical Products Corporation2 recalls0 reports
- Medline Industries, Inc. ECVC6370Centurion Medical Products Corporation · General Surgery Tray2 recalls0 reports
- MEDLINE ECVC6840Medline Industries, Inc. · General surgery tray2 recalls0 reports
- MEDLINE ECVC6265Medline Industries, Inc. · General Surgery Tray2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99729).
- Z-3230-2026 · Open, ClassifiedMedline convenience kits containing Healthmark Instrument Channel Brushes. SINUS PACK DYNJ48258D and SINUS PACK DYNJ48258F.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.