IntraSight Plus recallZ-3276-2026 · Volcano Corp · initiated Aug 6, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Volcano Corp or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Software system may associate image data from one patient with another patient's record, potentially resulting in incorrect diagnosis or treatment.

FDA root cause category: Software Design Change

What the company is doing

On August 6, 2026, Philips issued a Urgent Medical Device Correction to affected consignees. Philips ask consignees to take the following actions: 1. Contact your local Philips Service representative in advance of any changes to the worklist provider, Azurion software, interface configuration, or hospital IT environment, as such changes may affect the Study ID field. Your Philips Service representative will help evaluate any impact and/or reverify as applicable. Otherwise, you may continue to use the device in accordance with the intended use. 2. Please circulate this notice to all users of the IntraSight Plus so they are aware of the issue and associated hazard / harm. Philips also encourages all customers to post this letter on or near the system until Philips has updated its software.

Recall details

Recall number
Z-3276-2026
Status
Open, Classified
Initiated by the company
Aug 6, 2026
Posted by the FDA
Sep 25, 2026
Recalling firm
Volcano Corp
Quantity
19 systems
Product code
IYO
FDA submissions
K253714
Distribution
Worldwide - US Nationwide distribution in the states of MA, MN, TX, AL, GA, CA and the countries of Belgium, Denmark, Germany, Netherlands, Poland, Spain, Sweden, United Kingdom.

Product description

Philips IntraSight Plus REF: 797423 and 797417 100-120v-220-240v,50/60hZ, 1000 VA

Code information (models, lots, serial numbers)

UDI: REF: 797423:(01)00845225012946/REF:797417: (01)00845225012939 all Philips IntraSight Plus Systems with Software Version 6.0.4

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.