uCT recallZ-3280-2026 · Shanghai United Imaging Healthcare Co., Ltd. · initiated Aug 11, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Shanghai United Imaging Healthcare Co., Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
FDA root cause category: Software design
What the company is doing
On 8/11/2026, correction notices were emailed to customers informing them that software will be released to resolve the software problem. Customers were asked to do the following: When using the Study Split function, follow the requirements below: 1) Temporarily disable the "Study Split" function. Navigate to the system configuration interface and uncheck the "Auto Study Split" option. 2) If the "Study Split" function is clinically necessary, open the system configuration interface and select "Split on PACS" instead of "Split on Local". 3) When "Auto study split" is checked, ensure that PACS (Picture Archiving and Communication System) is configured on site and auto-archive is enabled. 4) If images are found to be missing on the PA interface, immediately verify whether the corresponding image data exists in the PACS. a) If images exist in PACS: They can be retrieved from PACS for review. b) If images do not exist in PACS: Contact the firm's service engineer for assistance. For further questions, contact firm service engineer or call firm Customer Contact Center at (855) 221-1552 or 832-699-2799
Recall details
- Recall number
- Z-3280-2026
- Status
- Open, Classified
- Initiated by the company
- Aug 11, 2026
- Posted by the FDA
- Sep 25, 2026
- Recalling firm
- Shanghai United Imaging Healthcare Co., Ltd.
- Quantity
- 14
- Product code
- JAK
- FDA submissions
- K172135, K230162, K241079, K253173
- Distribution
- US Nationwide distribution in the states of CA, PR, UT, WV, ND, NE, NC, TX, ID, LA, MT, FL, CO, IN, KY, AL, OH, MI, IA, MO, NY, NV, TN, WV, SC, WI, IL, MS, MA, NJ.
Product description
Computed Tomography X-ray System, Model: uCT 780 with software
Code information (models, lots, serial numbers)
UDI-DI: 06971576831050. Serial Number/Software: 684001/R001.17.6.1607336-Re3100SP4, 684002/R001.17.6.1607336-Re3100SP4, 684003/R003.8.4.1810621-Re3300SP1H1_P3, 684004/R003.8.4.1810621-Re3300SP1H1_P3, 684005/R003.8.4.1810621-Re3300SP1H1_P3, 684006/R003.8.4.1810621-Re3300SP1H1_P3, 684007/R003.8.4.1810621-Re3300SP1H1_P3, 684008/R003.8.4.1810621-Re3300SP1H1_P3, 684009/R003.8.4.1810621-Re3300SP1H1_P3, 684010/R003.8.4.1810621-Re3300SP1H1_P3, 684011/R003.8.4.1810621-Re3300SP1H1_P3, 684012/R003.9.0.1776134-Re-3300SP2, 684013/R003.11.0.1897081-Re3310SP1, 684014/R003.11.0.1897081-Re3310SP1
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- uCT uCT 780Shanghai United Imaging Healthcare Co., Ltd. · System, X-Ray, Tomography, Computed2 recalls0 reports
Product lines
Other devices under the same FDA submission
1 device cleared under K172135, K230162, K241079, K253173 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- uCT uCT 760Shanghai United Imaging Healthcare Co., Ltd. · System, X-Ray, Tomography, Computed2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99731).
- Z-3277-2026 · Open, ClassifiedComputed Tomography X-ray System, Model: uCT 530 with software
- Z-3278-2026 · Open, ClassifiedComputed Tomography X-ray System, Model: uCT 550 with software
- Z-3279-2026 · Open, ClassifiedComputed Tomography X-ray System, Model: uCT 760 with software
- Z-3281-2026 · Open, ClassifiedComputed Tomography X-ray System, Model: uCT ATLAS with software
- Z-3282-2026 · Open, ClassifiedComputed Tomography X-ray System, Model: uCT ATLAS Astound with software
- Z-3283-2026 · Open, ClassifiedPositron Emission Tomography and Computed Tomography System, Model: uMI 550 with software
- Z-3284-2026 · Open, ClassifiedPositron Emission Tomography and Computed Tomography System, Model: uMI Panorama 35 with software
- Z-3285-2026 · Open, ClassifiedPositron Emission Tomography and Computed Tomography System, Model: uMI Panorama GS with software
- Z-3286-2026 · Open, ClassifiedPositron Emission Tomography and Computed Tomography System, Model: uMI Panvivo with software
- Z-3287-2026 · Open, ClassifiedPositron Emission Tomography and Computed Tomography System, Model: uMI Panvivo S with software
- Z-3288-2026 · Open, ClassifiedPositron Emission Tomography and Computed Tomography System, Model: uMI 780
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page