MAS Omni IMMUNE PRO Liquid Assayed Integrated Immunoassay… recallZ-0159-2027 · Microgenics Corporation · initiated Aug 17, 2026
Class II Open, Classified All recalls posted in October 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Microgenics Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Vial-to-vial variability in MAS Omni"IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, Lot OPRO29061A results in low Parathyroid Hormone (PTH) recovery, which may lead to a delay in reporting patient PTH results.
FDA root cause category: Process change control
What the company is doing
On 8/17/2026, recall notices were emailed to customers asking them to do the following: 1) If the variability of the material (PTH Analyte) exceeds your laboratory's specification, dispose of the material. of the material in accordance with local, regional, and national regulations. 2) Contact [email protected] for any additional support. 3) Customers using the affected product can continue to use the product as long as it meets their laboratory's specification and continue monitoring performance of PTH analyte. 4) This recall notice needs to be passed to anyone within your organization or any organization where the potentially affected lot have been transferred. 5) Retain a copy of this letter for your laboratory records. 6) Complete and return the Recall Response Form via email to [email protected] In addition, distributors were asked to do the following: - Please notify your customers of this Recall Notice using this recall notice and request they return a response to your contact information. - Firm is providing a customer response form template as a word document. Please place this letter on to your own letterhead and collect responses from your customers on that customer response form. Please note, customers should provide their response directly to your contact information. If you have any additional questions regarding this Recall Notice, contact the firm at 1-800-232-3342 (USA and Canada only) or 510-979-5000 (International), option 2, then option 3 or by email: [email protected]
Recall details
- Recall number
- Z-0159-2027
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Aug 17, 2026
- Posted by the FDA
- Oct 15, 2026
- Recalling firm
- Microgenics Corporation
- Quantity
- 1366
- Product code
- JJY
- Distribution
- Worldwide - US Nationwide distribution in the states of UT, MI, OK, NY, TX, CA, LA, SC, OH, MO, OR and the countries of KOR, BGD, THA, ZAF, DEU, VNM, BRA, NZL.
Product description
MAS Omni IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, REF: OPRO-101; MAS Omni IMMUNETM PRO Liquid Assayed Integrated Immunoassay Control, Sample Pack, REF: OPRO-SP
Code information (models, lots, serial numbers)
REF/UDI-DI/Lot(Expiration): OPRO-101/00884883005327/OPRO29061A(2029-06-30); OPRO-SP/00884883005358/OPRO2906SA(2029-06-30)
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- MAS Omni-IMMUNE PRO Liquid Assayed Integrated Immunoassay Control OPRO-101Microgenics Corporation · Multi-Analyte Controls, All Kinds (Assayed)1 recall0 reports
- MAS Omni-IMMUNE PRO Liquid Assayed Integrated Immunoassay Control OPRO-SPMicrogenics Corporation · Multi-Analyte Controls, All Kinds (Assayed)1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Nov 27, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Nov 27, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Nov 27, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page