Skip to content

Brand Name: Opportunity Flex recallZ-3290-2026 · American Nitrile Operations Llc · initiated Aug 14, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from American Nitrile Operations Llc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Gloves may exhibit defects such as tears, missing fingers, and debris. Use of defective gloves may lead to healthcare personnel and patient exposure to infectious pathogens leading to infection or cross-contamination, or exposure to chemical agents leading to irritation or systemic toxicities.

FDA root cause category: Process control

What the company is doing

Firm began notifying its VA customers on August 15, 2026 via email. Instruction to Customers: 1. Immediately cease all use of the affected products. 2. Notify all users throughout your facility of the product recall. 3. Examine your inventory and immediately quarantine the affected lots to avoid accidental use 4 Contact distributor for replacement.

Recall details

Recall number
Z-3290-2026
Status
Open, Classified
Initiated by the company
Aug 14, 2026
Posted by the FDA
Sep 25, 2026
Recalling firm
American Nitrile Operations Llc
Quantity
36,793,000
Product code
LZA
FDA submissions
K220825
Distribution
US Nationwide distribution.

Product description

Brand Name: Opportunity Flex Product Name: Opportunity Flex Model/Catalog Number: 72-50900, 72-50901, 72-50902, 72-50903, 72-50904 Product Description: Blue Nitrile Patient examination gloves

Code information (models, lots, serial numbers)

Item No. 72-RM50900 UDI-DI 818634025517 Lots AO01292 AO01356 AO01430 AO01457 Item No. 72-RM50901 UDI-DI 818634025449 Lots AO01357 AO01431 AO01458 Item No. 72-RM50902 UDI-DI 818634025456 Lots AO01274 AO01351 AO01355 AO01426 AO01427 AO01429 AO01448 AO01456 AO01377 AO01384 Alternate lots AN250511407, AN250211237, AN250411331, AN250211252, AN250211236, AN250211237, AN250511407 Item No. 72-RM50903 UDI-DI 818634025463 Lots AO01309 AO01346 AO01353 AO01385 AO01396 AO01425 Alternate lot AN250511332 Item No. 72-RM50904 UDI-DI 818634025470 Lots AO01310 AO01339 AO01352 AO01354 AO01386 AO01408 Alternate lots AN250511409, AN250511333

Devices named in this recall

5 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 99748).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page