Medical convenience kits containing BD CliniClean Pericare… recallZ-3294-2026 · American Contract Systems Inc · initiated Aug 10, 2026
Class II Open, Classified All recalls posted in September 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from American Contract Systems Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Kits contain manufacturer-recalled perineal care (pericare) wipes within urinary catheter trays, which may be contaminated with a lubricant used during the manufacturing process. Use of a contaminated wipe may lead to localized skin effects such as irritation, erythema, or mild inflammation. Failure to perform pericare due to disposal of contaminated wipes may increase risk of localized infection or catheter-associated urinary tract infection (CAUTI).
FDA root cause category: Under Investigation by firm
What the company is doing
Firm began notifying affected distributors via email on August 10, 2026. Distributors were instructed to notify end-users. Customers were instructed to identify, segregate, and quarantine all impacted products. Customers are to add recall labels to affected kits to ensure the impacted component is discarded prior to use. An amended customer letter was sent out on August 28, 2026, which added the following instruction: Users should source alternative cleansing wipes for the impacted product lots and ensure they continue to follow their facility protocols for providing perineal care for patients with urinary catheters.
Recall details
- Recall number
- Z-3294-2026
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Aug 10, 2026
- Posted by the FDA
- Sep 25, 2026
- Recalling firm
- American Contract Systems Inc
- Quantity
- 1,804 kits total
- Product code
- OJF
- Distribution
- US Nationwide distribution in the states of Iowa, Minnesota, Wisconsin, and South Dakota.
Product description
Medical convenience kits containing BD CliniClean Pericare Wipes within BD Urinary Catheter Trays ABDOMINAL HYSTERECTOMY - 205989 Model No. MHAH55AM LAVH-LAP COLON Model No. SPLL54U VAG HYST PACK Model No. SPVH62P
Code information (models, lots, serial numbers)
MHAH55AM UDI-DI 191072236487 Lots 42-8529911 42-8649211 42-8687711 42-8797811 42-8857011 42-8869711 SPLL54U UDI-DI 191072239181 Lots 07-2516111 07-2528611 07-2545611 07-2547911 07-2561411 SPVH62P UDI-DI 191072238771 Lots 07-2544611 12-7436711
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- ACS SPVH62PAmerican Contract Systems, Inc. · Hysterectomy kit2 recalls0 reports
- ACS MHAH55AMAmerican Contract Systems, Inc. · Hysterectomy kit1 recall0 reports
- ACS SPLL54UAmerican Contract Systems, Inc. · Hysterectomy kit1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99759).
- Z-3292-2026 · Open, ClassifiedMedical convenience kits containing BD CliniClean Pericare Wipes within BD Urinary Catheter Trays ROBOTIC PACK - 242561 Model No. MHRB88AO
- Z-3293-2026 · Open, ClassifiedMedical convenience kits containing BD CliniClean Pericare Wipes within BD Urinary Catheter Trays CARDIAC HEART BUNDLE NS Model No. LMCH99P OPEN HEART 2 RCH 0248495 Model No.…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page