Halyard IV START KIT NON-STERILE recallZ-3261-2026 · AVID Medical, Inc. · initiated Jul 22, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from AVID Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Convenience kits containing BD ChloraPrep Clear - 1 mL Applicator, which may exhibit an open or incomplete seal on the packaging of the applicator.
FDA root cause category: Packaging process control
What the company is doing
Consignees were initially notified on about 07/22/2026 via "URGENT: MEDICAL DEVICE FIELD CORRECTION" letter. They were informed that product may exhibit an open or incomplete seal on the packaging of the applicator. Consignees were instructed to quarantine impacted product, discontinue use of the BD ChloraPrep Clear - 1 mL Applicators within the kits, add a warning label to instruct users to discard the impacted component, and notify customers if the kits had been further distributed. Consignees were notified again on about 08/20/2026 with an updated "URGENT: MEDICAL DEVICE FIELD CORRECTION" letter. Consignees were informed that certain lots of the BD ChloraPrep Applicators are contaminated with Aspergillus penicillioides. Instructions included to discard impacted kits containing affected applicators and notify customers if the kits had been further distributed. For those unaffected by the contamination, but are included in the action regarding sterility concerns, they were instructed to quarantine impacted product, discontinue use of the BD ChloraPrep Clear - 1 mL Applicators within the kits, add a warning label to instruct users to discard the impacted component, and notify customers if the kits had been further distributed.
Recall details
- Recall number
- Z-3261-2026
- Status
- Open, Classified
- Initiated by the company
- Jul 22, 2026
- Posted by the FDA
- Sep 23, 2026
- Recalling firm
- AVID Medical, Inc.
- Quantity
- 192 kits (4 cases)
- Product code
- LRO
- Distribution
- US Nationwide distribution.
Product description
Halyard IV START KIT NON-STERILE; 48 kits per case. Containing BD ChloraPrep Clear - 1 mL Applicator. Kit Model Number: KRIV26-01
Code information (models, lots, serial numbers)
Kit Model Number: KRIV26-01. UDI-DI: 10809160450109. Lot Number: 1642214. Expiration Date: 7/31/2027
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Halyard KRIV26-01_0001Avid Medical, Inc. · General surgery tray1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99764).
- Z-3260-2026 · Open, ClassifiedHalyard IV START KIT NON-STERILE; 48 kits per case. Containing BD ChloraPrep Clear - 1 mL Applicator. Kit Model Number: KRIV26-02
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page