Nalu TD2 Therapy Disc with firmware, REF: 34014-001, and… recallZ-0311-2027 · Boston Scientific Neuromodulation Corporation · initiated Sep 30, 2026
Class II Open, Classified All recalls posted in October 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Boston Scientific Neuromodulation Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Therapy Disc (TD) that powers and sends stimulation signals to an Implantable Pulse Generator (IPG), part of a neurostimulation system, has a firmware issue that causes the TD to attempt to restart, then fail, leading it to become non-functional, and if a secondary TD is unavailable, this may lead IPG to become non-functional, thus unable to inhibit transmission of pain signals.
FDA root cause category: Software change control
What the company is doing
On 9/30/2026, correction notices will be sent to healthcare professionals informing them of the following: Patients are provided with two TDs at the time of therapy activation, so therapy can be restored with the use of an alternate TD. The firm is asking healthcare professionals to do the following: 1) Identify the affected patients and inform them of this recall to arrange an-field TD firmware update. All affected units should be updated with the new firmware. Patient letter is provided with this notification for your use. 2) Contact the local firm representative to arrange an in-field TD firmware update at your facility. Append the patient's medical record accordingly when the firmware update is completed. 3) Share this communication with any healthcare professionals in your facility that use the product and with any other organization to which these systems may have been transferred. 4) Complete and return the Acknowledgement Form via email to [email protected] The patient notice recommended that patients contact their physician's office so they can arrange an in-field TD firmware update with the local firm representative. If they have any questions after contacting their doctor, contact the firm's Patient Care at 800-618-3402 or [email protected]. Report any adverse events or quality concerns associated with use of these devices to the firm via email at [email protected]
Recall details
- Recall number
- Z-0311-2027
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Sep 30, 2026
- Posted by the FDA
- Oct 29, 2026
- Recalling firm
- Boston Scientific Neuromodulation Corporation
- Quantity
- 134
- Product code
- GZB
- FDA submissions
- K233801
- Distribution
- US Nationwide distribution in the states of PA, FL, OR, AZ, OH, SC, TX, OR, OK, MI, CA, NJ.
Product description
Nalu TD2 Therapy Disc with firmware, REF: 34014-001, and TD2 Patient Kit, REF: 74011, part of the Nalu Neurostimulation System
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Firmware versions v2.0.71 and v2.0.77 REF/UDI-DI/Serial Numbers: 34014-001/00812537037134/ 100855, 100877, 100883, 100884, 100886, 100888, 100898, 100903, 100908, 100911, 100916, 100932, 101103, 100902; 74011/00812537037127, 00812537036977/101439, 100913, 100949, 101159, 101162, 101163, 101164, 101173, 101175, 101177, 101186, 101187, 101188, 101189, 101192, 101200, 101205, 101208, 101213, 101214, 101218, 101219, 101221, 101222, 101240, 101245, 101247, 101249, 101256, 101257, 101259, 101260, 101261, 101264, 101267, 101268, 101270, 101272, 101273, 101277, 101278, 101280, 101283, 101284, 101290, 101297, 101298, 101301, 101305, 101307, 101310, 101311, 101312, 101314, 101315, 101326, 101330, 101331, 101334, 101338, 101344, 101345, 101346, 101356, 101357, 101358, 101359, 101363, 101364, 101367, 101375, 101376, 101377, 101378, 101381, 101383, 101387, 101388, 101392, 101400, 101405, 101408, 101410, 101414, 101415, 101417, 101418, 101419, 101429, 101431, 101432, 101436, 101440, 101441, 101443, 101447, 101448, 101450, 101451, 101466, 101490, 101493, 101504, 101510, 101511, 101514, 101520, 101527, 101528, 101534, 101535, 101553, 101557, 101567, 101571, 101576, 101599, 101617, 101623, 101656
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- TD2 Patient Kit 74011Nalu Medical, Inc. · Stimulator, Peripheral Nerve, Implanted (Pain Relief)1 recall0 reports
- TD2 Therapy Disc 34014-001Nalu Medical, Inc. · Stimulator, Peripheral Nerve, Implanted (Pain Relief)1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Nov 27, 2026 · View the FDA recall notice ↗