Skip to content

Sperm Ab ELISA recallZ-0082-2027 · DRG International, Inc. · initiated Jul 2, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from DRG International, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for false negative patient results.

FDA root cause category: Reprocessing Controls

What the company is doing

On July 2, 2026 Field Safety Notice letters were sent to customers. Actions to be taken: Please check your stocks for the lot in question (320K046), stop using these kits immediately and quarantine these products without delay. Please destroy the kits and send us the confirmation with the completed response form. If you have transferred any of the affected products to another laboratory, please provide them with a copy of this letter.

Recall details

Recall number
Z-0082-2027
Status
Open, Classified
Initiated by the company
Jul 2, 2026
Posted by the FDA
Oct 2, 2026
Recalling firm
DRG International, Inc.
Quantity
211
Product code
DFQ
Distribution
Worldwide - US Nationwide distribution in the state of NJ and the countries of Azerbaijan, Belgium, Czech Republic, Germany, Hungary, India, Italy, Kuwait, Russia, Spain, Ukraine.

Product description

Sperm Ab ELISA; Model/Catalog Number: EIA-1826; Product Description: Sperm Ab ELISA (96 well plate test);

Code information (models, lots, serial numbers)

Model/Catalog Number: EIA-1826; UDI: 04048474018264; Lot# 320K046;

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page