NOxBOXi Nitric Oxide Delivery System. Model Number: NOXBOX-I recallZ-3306-2026 · Noxbox Ltd · initiated Sep 14, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Noxbox Ltd or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for irregular oxygen readings due to a manufacturing issue affecting certain oxygen sensors.
FDA root cause category: Component design/selection
What the company is doing
On about 08/31/2026 NOXBOX initially notified its single distributor of the recall via emailed URGENT MEDCIAL DEVICE CORRECTION letter. The letter instructed distributors to examine inventory for affected units, complete and return the provided Distributor Response Form, arrange for the replacement of affected oxygen sensors, perform all preventive maintenance checks on NOxBOXi devices prior to distribution of units, ensure all applicable personnel are notified within the organization, notify all customers who have received affected units. Device users were instructed to examine inventory for affected units, complete and return the provided Customer Response Form to their respective distributor, ensure all applicable personnel or organizations (if units were transferred) are notified within the organization, and ensure personnel using the device are made aware. NOxBOX is working with the oxygen sensor manufacturer to obtain replacement sensors for all affected devices and will coordinate with distributors to implement sensor replacement activities. As such, NOxBOX is asking distributors and customers to promptly return the Response Forms as instructed on the form to confirm the number devices still in service and the number of sensor replacements required.
Recall details
- Recall number
- Z-3306-2026
- Status
- Open, Classified
- Initiated by the company
- Sep 14, 2026
- Posted by the FDA
- Sep 25, 2026
- Recalling firm
- Noxbox Ltd
- Quantity
- 1,667 units
- Product code
- MRN
- FDA submissions
- K171696, K201339, K220898, K231823, K233251
- Distribution
- US Nationwide distribution in the state of Georgia.
Product description
NOxBOXi Nitric Oxide Delivery System. Model Number: NOXBOX-I.
Code information (models, lots, serial numbers)
Show the full code information (16k characters)
Model Number: NOXBOX-I. UDI-DI: 05060541640009. Serial Numbers: NI100825, NI100833, NI100839, NI100840, NI100841, NI100847, NI100848, NI100849, NI100850, NI100851, NI100852, NI100855, NI100858, NI100859, NI100861, NI100862, NI100865, NI100866, NI100868, NI100869, NI100872, NI100873, NI100875, NI100876, NI100878, NI100879, NI100880, NI100881, NI100882, NI100883, NI100884, NI100885, NI100886, NI100887, NI100888, NI100895, NI100932, NI100935, NI100942, NI100948, NI100949, NI100953, NI100961, NI100975, NI100976, NI100978, Serial Number, NI100980, NI100983, NI100985, NI100986, NI100987, NI100989, NI100990, NI100991, NI100993, NI100994, NI100996, NI100997, NI100999, NI101001, NI101002, NI101003, NI101004, NI101005, NI101006, NI101008, NI101009, NI101010, NI101011, NI101012, NI101013, NI101014, NI101015, NI101017, NI101018, NI101019, NI101024, NI101026, NI101028, NI101029, NI101030, NI101031, NI101033, NI101034, NI101035, NI101038, NI101039, NI101041, NI101042, NI101044, NI101048, NI101049, Serial Number, NI101052, NI101057, NI101059, NI101060, NI101061, NI101063, NI101065, NI101067, NI101069, NI101071, NI101072, NI101073, NI101077, NI101079, NI101080, NI101081, NI101083, NI101084, NI101086, NI101088, NI101089, NI101090, NI101091, NI101092, NI101093, NI101094, NI101095, NI101099, NI101100, NI101101, NI101105, NI101106, NI101107, NI101108, NI101109, NI101110, NI101111, NI101112, NI101113, NI101114, NI101115, NI101116, NI101117, NI101125, NI101130, NI101131, Serial Number, NI101132, NI101133, NI101134, NI101135, NI101136, NI101137, NI101138, NI101141, NI101142, NI101144, NI101145, NI101146, NI101147, NI101148, NI101149, NI101150, NI101151, NI101153, NI101154, NI101155, NI101156, NI101158, NI101159, NI101160, NI101161, NI101162, NI101163, NI101164, NI101165, NI101167, NI101168, NI101170, NI101173, NI101174, NI101175, NI101176, NI101177, NI101178, NI101179, NI101180, NI101181, NI101182, NI101183, NI101184, NI101190, NI101191, Serial Number, NI101192, NI101199, NI101202, NI101204, NI101205, NI101207, NI101208, NI101209, NI101211, NI101212, NI101213, NI101214, NI101215, NI101216, NI101217, NI101218, NI101219, NI101220, NI101221, NI101222, NI101223, NI101224, NI101225, NI101227, NI101228, NI101229, NI101230, NI101231, NI101232, NI101233, NI101234, NI101236, NI101237, NI101241, NI101243, NI101244, NI101245, NI101246, NI101249, NI101250, NI101251, NI101252, NI101253, NI101254, NI101255, NI101256, Serial Number, NI101257, NI101258, NI101259, NI101263, NI101264, NI101265, NI101266, NI101268, NI101270, NI101271, NI101273, NI101278, NI101282, NI101284, NI101285, NI101286, NI101287, NI101288, NI101289, NI101292, NI101294, NI101295, NI101297, NI101298, NI101299, NI101300, NI101301, NI101303, NI101305, NI101307, NI101308, NI101309, NI101313, NI101314, NI101317, NI101321, NI101322, NI101323, NI101324, NI101326, NI101327, NI101328, NI101329, NI101330, NI101331, NI101333, Serial Number, NI101335, NI101336, NI101337, NI101338, NI101340, NI101342, NI101344, NI101346, NI101347, NI101349, NI101350, NI101351, NI101353, NI101354, NI101355, NI101356, NI101357, NI101359, NI101360, NI101361, NI101368, NI101372, NI101373, NI101376, NI101377, NI101379, NI101380, NI101381, NI101384, NI101386, NI101387, NI101388, NI101390, NI101392, NI101393, NI101394, NI101395, NI101396, NI101402, NI101404, NI101405, NI101406, NI101407, NI101408, NI101409, NI101410, Serial Number, NI101411, NI101412, NI101413, NI101414, NI101415, NI101416, NI101423, NI101425, NI101426, NI101428, NI101429, NI101430, NI101431, NI101432, NI101433, NI101434, NI101435, NI101436, NI101437, NI101439, NI101441, NI101442, NI101443, NI101444, NI101446, NI101448, NI101449, NI101451, NI101452, NI101454, NI101456, NI101457, NI101459, NI101460, NI101461, NI101462, NI101463, NI101464, NI101467, NI101468, NI101469, NI101472, NI101473, NI101474, NI101475, NI101478, Serial Number, NI101479, NI101482, NI101483, NI101484, NI101485, NI101486, NI101487, NI101489, NI101490, NI101491, NI101492, NI101496, NI101505, NI101506, NI101508, NI101509, NI101512, NI101514, NI101515, NI101516, NI101520, NI101521, NI101522, NI101524, NI101525, NI101526, NI101527, NI101529, NI101530, NI101532, NI101533, NI101540, NI101550, NI101551, NI101552, NI101553, NI101554, NI101557, NI101560, NI101563, NI101565, NI101566, NI101569, NI101570, NI101572, NI101573, Serial Number, NI101574, NI101575, NI101576, NI101577, NI101578, NI101579, NI101581, NI101584, NI101585, NI101588, NI101590, NI101591, NI101592, NI101593, NI101594, NI101595, NI101596, NI101597, NI101598, NI101599, NI101600, NI101601, NI101602, NI101604, NI101605, NI101606, NI101607, NI101608, NI101609, NI101610, NI101611, NI101612, NI101613, NI101614, NI101615, NI101616, NI101617, NI101620, NI101622, NI101623, NI101624, NI101625, NI101626, NI101627, NI101628, NI101629, Serial Number, NI101635, NI101636, NI101638, NI101639, NI101640, NI101641, NI101643, NI101644, NI101645, NI101651, NI101654, NI101655, NI101656, NI101657, NI101658, NI101659, NI101660, NI101662, NI101664, NI101665, NI101666, NI101667, NI101668, NI101669, NI101670, NI101671, NI101672, NI101674, NI101675, NI101676, NI101677, NI101678, NI101679, NI101680, NI101681, NI101682, NI101683, NI101684, NI101685, NI101686, NI101687, NI101688, NI101689, NI101690, NI101691, NI101692, Serial Number, NI101693, NI101694, NI101695, NI101696, NI101697, NI101699, NI101700, NI101703, NI101704, NI101705, NI101706, NI101707, NI101708, NI101709, NI101710, NI101711, NI101712, NI101713, NI101714, NI101715, NI101716, NI101717, NI101718, NI101719, NI101720, NI101721, NI101722, NI101723, NI101724, NI101726, NI101727, NI101729, NI101731, NI101732, NI101733, NI101735, NI101736, NI101737, NI101739, NI101740, NI101742, NI101743, NI101744, NI101746, NI101747, NI101748, Serial Number, NI101749, NI101750, NI101751, NI101753, NI101756, NI101759, NI101760, NI101761, NI101763, NI101765, NI101766, NI101767, NI101768, NI101769, NI101770, NI101771, NI101775, NI101776, NI101777, NI101778, NI101781, NI101782, NI101783, NI101784, NI101787, NI101791, NI101793, NI101794, NI101795, NI101796, NI101797, NI101798, NI101799, NI101804, NI101805, NI101806, NI101807, NI101808, NI101809, NI101810, NI101811, NI101812, NI101813, NI101814, NI101815, NI101816, Serial Number, NI101817, NI101818, NI101819, NI101820, NI101821, NI101822, NI101823, NI101824, NI101825, NI101826, NI101828, NI101829, NI101831, NI101832, NI101834, NI101835, NI101836, NI101837, NI101838, NI101840, NI101841, NI101842, NI101843, NI101844, NI101845, NI101846, NI101848, NI101849, NI101851, NI101854, NI101857, NI101859, NI101861, NI101862, NI101864, NI101865, NI101867, NI101868, NI101869, NI101872, NI101873, NI101874, NI101875, NI101876, NI101878, NI101879, Serial Number, NI101880, NI101881, NI101882, NI101883, NI101884, NI101885, NI101886, NI101887, NI101889, NI101890, NI101891, NI101892, NI101895, NI101896, NI101898, NI101899, NI101901, NI101902, NI101903, NI101904, NI101905, NI101906, NI101907, NI101908, NI101909, NI101910, NI101912, NI101913, NI101914, NI101916, NI101919, NI101920, NI101922, NI101924, NI101925, NI101926, NI101928, NI101936, NI101937, NI101939, NI101940, NI101941, NI101942, NI101943, NI101944, NI101945, Serial Number, NI101946, NI101947, NI101948, NI101952, NI101953, NI101954, NI101955, NI101956, NI101957, NI101958, NI101959, NI101960, NI101961, NI101962, NI101963, NI101964, NI101965, NI101966, NI101967, NI101968, NI101970, NI101971, NI101972, NI101973, NI101974, NI101975, NI101976, NI101977, NI101978, NI101979, NI101981, NI101982, NI101983, NI101984, NI101985, NI101986, NI101987, NI101988, NI101989, NI101990, NI101991, NI101992, NI101993, NI101994, NI101995, NI101996, Serial Number, NI101997, NI101998, NI101999, NI102000, NI102001, NI102002, NI102003, NI102004, NI102006, NI102007, NI102008, NI102009, NI102010, NI102011, NI102012, NI102013, NI102014, NI102015, NI102017, NI102018, NI102019, NI102020, NI102021, NI102023, NI102024, NI102025, NI102026, NI102027, NI102028, NI102031, NI102032, NI102033, NI102035, NI102036, NI102037, NI102038, NI102040, NI102043, NI102044, NI102045, NI102046, NI102047, NI102048, NI102049, NI102051, NI102052, Serial Number, NI102053, NI102055, NI102056, NI102057, NI102059, NI102060, NI102061, NI102062, NI102063, NI102065, NI102067, NI102068, NI102069, NI102070, NI102071, NI102072, NI102074, NI102075, NI102076, NI102077, NI102078, NI102079, NI102081, NI102082, NI102083, NI102086, NI102089, NI102090, NI102091, NI102092, NI102093, NI102094, NI102095, NI102096, NI102097, NI102098, NI102099, NI102102, NI102103, NI102104, NI102105, NI102106, NI102107, NI102108, NI102109, NI102110, Serial Number, NI102111, NI102113, NI102114, NI102115, NI102116, NI102117, NI102118, NI102119, NI102120, NI102121, NI102123, NI102124, NI102125, NI102126, NI102127, NI102128, NI102129, NI102131, NI102132, NI102133, NI102134, NI102135, NI102136, NI102137, NI102138, NI102139, NI102140, NI102141, NI102142, NI102144, NI102145, NI102146, NI102147, NI102148, NI102149, NI102150, NI102152, NI102155, NI102160, NI102161, NI102163, NI102164, NI102165, NI102166, NI102167, NI102168, Serial Number, NI102169, NI102170, NI102171, NI102172, NI102173, NI102174, NI102175, NI102176, NI102177, NI102180, NI102182, NI102183, NI102184, NI102186, NI102188, NI102189, NI102190, NI102191, NI102192, NI102193, NI102194, NI102195, NI102196, NI102197, NI102198, NI102199, NI102200, NI102205, NI102206, NI102208, NI102209, NI102212, NI102213, NI102214, NI102215, NI102216, NI102219, NI102221, NI102222, NI102226, NI102227, NI102228, NI102230, NI102231, NI102232, NI102233, Serial Number, NI102234, NI102235, NI102237, NI102238, NI102239, NI102240, NI102241, NI102244, NI102245, NI102246, NI102247, NI102249, NI102250, NI102253, NI102254, NI102255, NI102256, NI102259, NI102263, NI102265, NI102266, NI102268, NI102269, NI102270, NI102271, NI102272, NI102273, NI102274, NI102275, NI102276, NI102277, NI102278, NI102279, NI102280, NI102281, NI102282, NI102283, NI102284, NI102285, NI102286, NI102287, NI102288, NI102290, NI102291, NI102292, NI102293, Serial Number, NI102295, NI102296, NI102297, NI102298, NI102299, NI102300, NI102304, NI102305, NI102306, NI102307, NI102308, NI102309, NI102310, NI102311, NI102327, NI102329, NI102330, NI102331, NI102332, NI102338, NI102339, NI102340, NI102341, NI102342, NI102343, NI102345, NI102346, NI102347, NI102349, NI102350, NI102351, NI102355, NI102356, NI102357, NI102358, NI102359, NI102360, NI102361, NI102362, NI102363, NI102364, NI102366, NI102377, NI102380, NI102381, NI102383, Serial Number, NI102384, NI102385, NI102387, NI102389, NI102390, NI102391, NI102392, NI102393, NI102394, NI102395, NI102396, NI102397, NI102398, NI102399, NI102400, NI102402, NI102403, NI102404, NI102405, NI102406, NI102407, NI102408, NI102410, NI102411, NI102413, NI102414, NI102415, NI102417, NI102418, NI102422, NI102423, NI102424, NI102425, NI102426, NI102427, NI102428, NI102429, NI102431, NI102432, NI102433, NI102434, NI102435, NI102436, NI102437, NI102438, NI102444, Serial Number, NI102445, NI102447, NI102463, NI102466, NI102467, NI102472, NI102473, NI102474, NI102475, NI102476, NI102477, NI102478, NI102479, NI102480, NI102481, NI102482, NI102483, NI102484, NI102485, NI102486, NI102487, NI102488, NI102489, NI102490, NI102492, NI102494, NI102495, NI102496, NI102497, NI102498, NI102499, NI102500, NI102501, NI102502, NI102503, NI102504, NI102506, NI102507, NI102509, NI102510, NI102511, NI102512, NI102513, NI102517, NI102518, NI102519, Serial Number, NI102520, NI102522, NI102525, NI102528, NI102530, NI102531, NI102535, NI102536, NI102537, NI102538, NI102539, NI102540, NI102541, NI102542, NI102543, NI102544, NI102545, NI102546, NI102547, NI102548, NI102550, NI102551, NI102552, NI102553, NI102554, NI102555, NI102556, NI102557, NI102558, NI102560, NI102561, NI102562, NI102563, NI102564, NI102565, NI102566, NI102567, NI102568, NI102569, NI102570, NI102571, NI102574, NI102575, NI102576, NI102577, NI102578, Serial Number, NI102579, NI102580, NI102581, NI102582, NI102583, NI102584, NI102585, NI102586, NI102587, NI102588, NI102590, NI102591, NI102592, NI102593, NI102594, NI102595, NI102596, NI102597, NI102598, NI102599, NI102600, NI102601, NI102602, NI102604, NI102605, NI102606, NI102607, NI102608, NI102610, NI102611, NI102613, NI102614, NI102615, NI102616, NI102617, NI102618, NI102619, NI102620, NI102621, NI102622, NI102623, NI102624, NI102625, NI102627, NI102628, NI102629, Serial Number, NI102630, NI102631, NI102632, NI102634, NI102635, NI102637, NI102638, NI102639, NI102640, NI102641, NI102642, NI102643, NI102645, NI102647, NI102648, NI102649, NI102651, NI102652, NI102653, NI102655, NI102656, NI102657, NI102658, NI102659, NI102660, NI102661, NI102662, NI102663, NI102664, NI102665, NI102666, NI102667, NI102668, NI102669, NI102670, NI102671, NI102672, NI102673, NI102674, NI102675, NI102676, NI102680, NI102681, NI102683, NI102684, NI102685, Serial Number, NI102686, NI102687, NI102688, NI102689, NI102691, NI102692, NI102693, NI102695, NI102696, NI102697, NI102698, NI102699, NI102700, NI102701, NI102702, NI102703, NI102704, NI102705, NI102706, NI102707, NI102708, NI102709, NI102710, NI102711, NI102713, NI102714, NI102715, NI102717, NI102720, NI102722, NI102724, NI102725, NI102726, NI102727, NI102728, NI102729, NI102730, NI102731, NI102732, NI102733, NI102734, NI102735, NI102736, NI102737, NI102738, NI102739, Serial Number, NI102740, NI102741, NI102742, NI102743, NI102744, NI102745, NI102746, NI102747, NI102748, NI102751, NI102752, NI102753, NI102755, NI102756, NI102757, NI102758, NI102759, NI102760, NI102762, NI102763, NI102764, NI102765, NI102766, NI102767, NI102768, NI102769, NI102770, NI102771, NI102772, NI102773, NI102774, NI102776, NI102777, NI102778, NI102781, NI102783, NI102784, NI102785, NI102786, NI102789, NI102790, NI102791, NI102792, NI102793, NI102794, NI102795, Serial Number, NI102798, NI102799, NI102800, NI102801, NI102802, NI102803, NI102804, NI102805, NI102806, NI102807, NI102808, NI102809, NI102810, NI102811, NI102812, NI102813, NI102814, NI102815, NI102816, NI102817, NI102818, NI102819, NI102821, NI102823, NI102824, NI102825, NI102828, NI102830, NI102832, NI102833, NI102834, NI102836, NI102837, NI102838, NI102839, NI102840, NI102841, NI102842, NI102843, NI102844, NI102845, NI102846, NI102847, NI102848, NI102849, NI102850, Serial Number, NI102851, NI102852, NI102853, NI102854, NI102855, NI102856, NI102858, NI102859, NI102860, NI102861, NI102862, NI102863, NI102865, NI102866, NI102868, NI102870, NI102871, NI102872, NI102873, NI102874, NI102876, NI102877, NI102878, NI102879, NI102880, NI102881, NI102882, NI102883, NI102884, NI102885, NI102886, NI102887, NI102888, NI102889, NI102890, NI102891, NI102892, NI102893, NI102894, NI102897, NI102898, NI102899, NI102900, NI102901, NI102902, NI102904, Serial Number, NI102905, NI102906, NI102907, NI102908, NI102910, NI102911, NI102913, NI102914, NI102915, NI102916, NI102917, NI102918, NI102919, NI102920, NI102921, NI102922, NI102923, NI102926, NI102927, NI102928, NI102929, NI102930, NI102931, NI102932, NI102933, NI102934, NI102935, NI102936, NI102937, NI102939, NI102942, NI102944, NI102947, NI102950, NI102955, NI102957, NI102958, NI102961, NI102962, NI102963, NI102964, NI102966, NI102967, NI102968, NI102970, NI102973, Serial Number, NI102977, NI102979, NI102986, NI102987, NI102989, NI102990, NI102991, NI102992, NI102996, NI102997, NI102998, NI102999, NI103001, NI103002, NI103003, NI103007, NI103011, NI103012, NI103014, NI103016, NI103017, NI103018, NI103019, NI103020, NI103021, NI103022, NI103023, NI103024, NI103074, NI103079, NI103082, NI103083, NI103084, NI103085, NI103087, NI103088, NI103089, NI103091, NI103093, NI103094, NI103095, NI103104, NI103116, NI103117, NI103120, NI103122, Serial Number, NI103123, NI103126, NI103127, NI103131, NI103136, NI103139, NI103144, NI103145, NI103146, NI103147, NI103148, NI103151, NI103152, NI103153, NI103155, NI103157, NI103159, NI103168, NI103170, NI103181, NI103182, NI103183, NI103189, NI103196, NI103203, NI103204, NI103208, NI103215, NI103220. Includes NOxBOXi with Oxygen sensor lots manufactured before April 13, 2026
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- NOxBOXi NOXBOX-INoxbox Ltd · Analyzer, Gas, Oxygen, Gaseous-Phase6 recalls5 deaths121 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page