Skip to content

Intellijoint Navigation System: Workstation G2 Configured… recallZ-0119-2027 · Intellijoint Surgical, Inc. · initiated Aug 28, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Intellijoint Surgical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A navigation system used to aid in selection and positioning of orthopedic implant system components has a workstation that may contain a 15-inch MacBook Pro battery, subject to an Apple battery recall, that may overheat and pose a fire safety risk.

FDA root cause category: Nonconforming Material/Component

What the company is doing

On 8/28/2026, recall notices were emailed to customers who were asked to do the following: - If affected serial numbers are found, stop use and contact firm's customer service: at [email protected] and copy [email protected] - If the observed serial number does not match the affected serial number please record the serial number identified and inform [email protected] and cc [email protected] of the correct serial number for that facility.

Recall details

Recall number
Z-0119-2027
Status
Open, Classified
Initiated by the company
Aug 28, 2026
Posted by the FDA
Oct 7, 2026
Recalling firm
Intellijoint Surgical, Inc.
Quantity
8
Product code
OLO
FDA submissions
K191507
Distribution
Worldwide - US Nationwide distribution in the states of CA, TX, MS, NJ and the countries of Singapore, Spain.

Product description

Intellijoint Navigation System: Workstation G2 Configured, REF: 000-0241; Intellijoint HIP Cart System, REF: 000-0250; Intellijoint HIP, Portable System, REF: 000-0249

Code information (models, lots, serial numbers)

REF/UDI-DI/MacBook Serial Numbers: 000-0241/00628184002414/C02RP36YG8WP, C02QPCF5G8WN 000-0250/00628184002506/C02RF44TG8WN, C02QW25RG8WL, C02QH355G8WM, C02S22N9G8WP; 000-0249/00628184002490/C02RQ3FUG8WN, C02R4451G8WN

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates