Skip to content

Panorama and Telemetry Monitoring System. Operator's… recallZ-0112-2027 · Mindray DS USA, Inc. · initiated Aug 31, 2026

Open, Classified All recalls posted in October 2026

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Mindray DS USA, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Certain monitor operator's manuals contain descriptions that may be insufficient or may require clarification regarding selected physiological alarm functions

FDA root cause category: Labeling design

What the company is doing

Mindray notified consignees on about 08/31/2026 via "URGENT: Medical Device Recall" letter. Consignees were informed that Mindray was providing updated instructions regarding physiological alarms in certain Mindray patient monitoring device Instructions for Use (IFU)/Operator s Manuals. This action is limited to updates to the IFU/Operators Manual content and does not require removal, return, repair, or replacement of the monitors. Consignees were instructed to continue using the Mindray monitors associated with the affected operator s manual, ensure that applicable clinical staff are made aware of the content of this letter and the Operator s Manual Addendum, and if monitors have been distributed or transferred to another organization, please forward this notice to that organization. Additionally, consignees were instructed to complete and return the Acknowledgment and Receipt Form provided. Distributors were instructed to notify their customers. Mindray provided the QR Code for the Operator s Manual Addendum - Addendum of the Arrhythmia Analysis Statement on their portal.

Recall details

Recall number
Z-0112-2027
Recall class
Not yet classified by the FDA
Status
Open, Classified
Initiated by the company
Aug 31, 2026
Posted by the FDA
Oct 7, 2026
Recalling firm
Mindray DS USA, Inc.
Quantity
11,831 units
Product code
MSX
FDA submissions
K242728
Distribution
Worldwide - Nationwide and the country of Canada.

Product description

Panorama and Telemetry Monitoring System. Operator's Manuals: TMS-6016/Panorama Telepack-608/TMS60 (TD60).

Code information (models, lots, serial numbers)

Part Numbers of the Panorama and Telemetry Monitoring System Operator's Manuals: 1. Panorama Telepack-608 Telemetry Monitoring System Operator's Manual up to and including revision 1.0; Part No. 046-007526-00 2. TMS-6016 Operator's Manual up to and including revision 11.0; Part No. 046-005012-00

Other devices under the same FDA submission

1 device cleared under K242728 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 99828).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page