Why it was recalled
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
FDA root cause category: Under Investigation by firm
What the company is doing
On 08/24/2026, an "URGENT MEDICAL DEVICE RECALL" Letter was emailed to customers informing them that, ASP Global identified a discrepancy between the labeled shelf life and available shelf-life validation data for certain Sterile Disposable Tourniquet Cuff products. While affected products were labeled with a 10-year shelf life, available validation documentation supports a shelf life of 5-years. Customer are instructed to: a) Review Appendix A to identify affected item codes and lot numbers. b) Discontinue use of the affected products. c) Quarantine any remaining impacted inventory to prevent further use. d) Complete and return the enclosed Recall Response Form to confirm receipt and actions to [email protected]. e) Dispose remaining inventory in accordance with the local facility s guidelines or notify ASP Global at [email protected] for return authorization and disposal assistance. f) If customers have distributed affected products to other facilities or customers, notify those recipients of this recall and provide them with a copy of this notification. For questions regarding this recall, contact: ASP Global Field Actions Team Email: [email protected] Phone: (855)-867-2190| Mon Fri, 9:00 AM 5:00 PM EST
Recall details
- Recall number
- Z-0044-2027
- Status
- Open, Classified
- Initiated by the company
- Aug 24, 2026
- Posted by the FDA
- Oct 1, 2026
- Recalling firm
- ASP Global, LLC.
- Quantity
- 45960
- Product code
- KCY
- Distribution
- US Nationwide distribution in the states of MO, NV, IN, MN, GA, WI, PA, TX, CT, LA, NJ, FL, AL, NY, SC, NE, NC, OH, AZ, CO, MA, IL, KY, CA, VA, ME, KS, MD, TN, MS, AR, WA, NM, MI, WY.
Product description
Brand Name: ASP Global Product Name:DISPOSABLE TOURNIQUET CUFF, 30" X 4" 2-Line, White Model/Catalog Number: ASP-564011 Software Version: Not Applicable Product Description: Sterile Disposable Tourniquet Cuff is used to be wrapped around a patient's limb and inflated to reduce or totally occlude circulation during surgery. It is non-invasive device. Tourniquet Cuffs are intended for use in surgical procedures that requires the interruption of blood flow and clear visualization of the operative field. Component: No
Code information (models, lots, serial numbers)
Lot Code: Model Number: ASP-564011 UDI-DI: (01) 00840216106765 All Lots produced from 01SEP2016 to 01MAR2023: P1015STY, P1215STY, O0815STY, O1020STY, N0330STY, N0530STY, M0430STY, M0730STY, M1015STY, M1101STY, M1115STY, L0228STY, L0310STY, L0325STY
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- ASP MEDICAL ASP-564011Asp Global, LLC · Tourniquet, Pneumatic1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99676).
- Z-0038-2027 · Open, ClassifiedBrand Name: ASP Global Product Name:DISPOSABLE TOURNIQUET CUFF, 12" X 3.5" 2-Line, Plum Model/Catalog Number: ASP-564003 Software Version: Not Applicable Product Description: S…
- Z-0039-2027 · Open, ClassifiedBrand Name: ASP Global Product Name:DISPOSABLE TOURNIQUET CUFF, 18" X 4" 1-Line, Red Model/Catalog Number: ASP-564006 Software Version: Not Applicable Product Description: Ster…
- Z-0040-2027 · Open, ClassifiedBrand Name: ASP Global Product Name:DISPOSABLE TOURNIQUET CUFF, 18" X 4" 2-Line, Red Model/Catalog Number: ASP-564007 Software Version: Not Applicable Product Description: Ster…
- Z-0041-2027 · Open, ClassifiedBrand Name: ASP Global Product Name:DISPOSABLE TOURNIQUET CUFF, 24" X 4" 1-Line, Yellow Model/Catalog Number: ASP-564008 Software Version: Not Applicable Product Description: S…
- Z-0042-2027 · Open, ClassifiedBrand Name: ASP Global Product Name:DISPOSABLE TOURNIQUET CUFF, 24" X 4" 2-Line, Yellow Model/Catalog Number: ASP-564009 Software Version: Not Applicable Product Description: S…
- Z-0043-2027 · Open, ClassifiedBrand Name: ASP Global Product Name:DISPOSABLE TOURNIQUET CUFF, 30" X 4" 1-Line, White Model/Catalog Number: ASP-564010 Software Version: Not Applicable Product Description: St…
- Z-0045-2027 · Open, ClassifiedBrand Name: ASP Global Product Name:DISPOSABLE TOURNIQUET CUFF, 34" X 4" 1-Line, Blue Model/Catalog Number: ASP-564012 Software Version: Not Applicable Product Description: S…
- Z-0046-2027 · Open, ClassifiedBrand Name: ASP Global Product Name:DISPOSABLE TOURNIQUET CUFF, 34" X 4" 2-Line, Blue Model/Catalog Number: ASP-564013 Software Version: Not Applicable Product Description: Ste…
- Z-0047-2027 · Open, ClassifiedBrand Name: ASP Global Product Name:DISPOSABLE TOURNIQUET CUFF, 44" X 4" 1-Line, Green Model/Catalog Number: ASP-564014 Software Version: Not Applicable Product Description: St…
- Z-0048-2027 · Open, ClassifiedBrand Name: ASP Global Product Name:DISPOSABLE TOURNIQUET CUFF, 44" X 4" 2-Line, Green Model/Catalog Number: ASP-564015 Software Version: Not Applicable Product Description: …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page