Skip to content

Precision Pointer Electromagnetic (8000-050-001) recallZ-0314-2027 · Stryker Leibinger GmbH & Co. KG · initiated Sep 24, 2026

Class II Open, Classified All recalls posted in October 2026

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Stryker Leibinger GmbH & Co. KG or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Device may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the device is inside the patient.

FDA root cause category: Process control

What the company is doing

The firm began notifying customers on September 24, 2026 via Urgent Medical Device Recall letters. Customers were instructed to immediately locate and quarantine any affected products at their facility. Customers may either destroy affected products onsite with their Stryker representative and send a completed Certificate of Destruction to the firm, or return the products to Stryker. The firm will provide a replacement upon receipt of the recalled product or Certificate of Destruction.

Recall details

Recall number
Z-0314-2027
Recall class
Class II
Status
Open, Classified
Initiated by the company
Sep 24, 2026
Posted by the FDA
Oct 29, 2026
Recalling firm
Stryker Leibinger GmbH & Co. KG
Quantity
513
Product code
HAW
FDA submissions
K212194
Distribution
Worldwide - US Nationwide distribution in the states of Alabama, Arizona, California, Colorado, Connecticut, Delaware, Florida, Hawaii, Illinois, Indiana, Kentucky, Louisiana, Massachusetts, Michigan, Missouri, Montana, New Hampshire, New York, Ohio, Pennsylvania, South Carolina, South Dakota, Tennessee, Texas, Utah, Washington and the countries of Austria, Canada, Denmark, Finland, France, Germany, Ireland, Italy, Japan, Norway, Portugal, South Korea, Spain, Switzerland, UK.

Product description

Precision Pointer Electromagnetic (8000-050-001)

Code information (models, lots, serial numbers)

Lot 7000413361 UDI-DI 07613327413205

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page