Williams recallZ-0163-2027 · Cook Incorporated · initiated Sep 16, 2026
Class II Open, Classified All recalls posted in October 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Cook Incorporated or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The female Luer lock adapter component in the affected device lots may not have been correctly seated due to misalignment and inadequate torquing during manufacturing. Potential events that may occur as a result of this issue include increased procedural time and inadequate treatment.
FDA root cause category: Under Investigation by firm
What the company is doing
An URGENT: MEDICAL DEVICE RECALL notification letter dated 9/16/26 was sent to customers. Actions to be Taken by the Customer 1. Examine inventory immediately to determine if you have affected product(s) and quarantine any affected product that remains unused. Immediately cease all further distribution and use of the affected products. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Refer to the Acknowledgement and Receipt Form for return instructions. 3. Complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter regardless of whether you have affected product(s) on hand. You may return the Form via fax (812.339.7316) or email ([email protected]). 4. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the affected devices have been transferred. 5. Please immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: [email protected]. Action being Taken by Cook Medical Cook Medical is removing potentially affected products from the market. Additionally, an investigation has been opened to determine the appropriate actions to prevent reoccurrence of a similar issue. If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235.
Recall details
- Recall number
- Z-0163-2027
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Sep 16, 2026
- Posted by the FDA
- Oct 15, 2026
- Recalling firm
- Cook Incorporated
- Quantity
- 97 units
- Product code
- FBK
- FDA submissions
- K171602
- Distribution
- US States: FL, IL, NM, TX, WA. Canada, France, Germany, Italy, Spain, Switzerland.
Product description
Williams Cystoscopic Injection Needles, used for cystoscopic injection into the urethra, bladder neck, and bladder wall, Reference Part Number 090001-S26, Order Number (GPN) G16112
Code information (models, lots, serial numbers)
UDI-DI: 00827002161122; Lot Numbers: 16613606, 16951548, 16951548X
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Williams G16112Cook Incorporated · endoscopic injection needle, gastroenterology-urology1 recall0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Nov 27, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Nov 27, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Nov 27, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page