INSTRUMENT recallZ-0166-2027 · Howmedica Osteonics Corp. · initiated Sep 16, 2026
Class II Open, Classified All recalls posted in October 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Howmedica Osteonics Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
FDA root cause category: Labeling design
What the company is doing
Stryker notified consignees on about 09/16/2026 via "MEDICAL DEVICE CORRECTION" letter mailed through UPS 2-day air. Consignees were informed that reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes, however, these trays require a sterilization dry time of 40 minutes when sterilized in blue wrap. Consignees were instructed to inform users of the Medical Device Correction and forward the notice to all individuals who need to be made aware or organizations who have consigned product, review and follow the updated instructions provided in LSTPI-B Cleaning, Sterilization, Inspection, and Maintenance of Reusable Medical Devices Reference Guide, complete and return the provided Medical Device Correction Business Reply Form.
Recall details
- Recall number
- Z-0166-2027
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Sep 16, 2026
- Posted by the FDA
- Oct 15, 2026
- Recalling firm
- Howmedica Osteonics Corp.
- Quantity
- 359 trays
- Product code
- FSM
- Distribution
- These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS.
Product description
Triathlon Hinge Femoral Prep Tray. Catalog Number: 6543-6-514.
Code information (models, lots, serial numbers)
Catalog Number: 6543-6-514. GTIN: 07613327562859. All lot numbers used prior to the correction date of May 8, 2026 are impacted.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- INSTRUMENT 6543-6-514Howmedica Osteonics Corp. · TRAY, SURGICAL, INSTRUMENT1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 99883).
- Z-0164-2027 · Open, ClassifiedTS Universal Baseplate Finishing Tray, Triathlon. Catalog Number: 6543-6-506.
- Z-0165-2027 · Open, ClassifiedTS Femoral Prep Tray, Triathlon. Catalog Number: 6543-6-507.
- Z-0167-2027 · Open, ClassifiedTS Misc. Instruments Tray, Triathlon. Catalog Number: 6543-6-519.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Nov 27, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Nov 27, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Nov 27, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page