Ion Torrent Genexus Dx with software, Catalog: A40003539… recallZ-0211-2027 · Life Technologies Holdings Pte Ltd · initiated Aug 14, 2026
Class II Open, Classified All recalls posted in October 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Life Technologies Holdings Pte Ltd or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
DNA or RNA sequencing system software has vulnerabilities.
FDA root cause category: Software design
What the company is doing
On 8/14/2026, recall notices were emailed to customers who were informed of the following: Once the software becomes available, your field support specialist or your local Service and Support representative will reach out to schedule a software upgrade. Firm recommends that customers back up their data before any software upgrades. Please refer to "Manage backup settings and restore runs (administrator)" section of the user guide. Complete and return the CUSTOMER RESPONSE Acknowledgement and Receipt Form via email to [email protected]; and also contact the firm at this email address is you have any technical questions or concerns.
Recall details
- Recall number
- Z-0211-2027
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Aug 14, 2026
- Posted by the FDA
- Oct 22, 2026
- Recalling firm
- Life Technologies Holdings Pte Ltd
- Quantity
- 16
- Product code
- PFF
- Distribution
- US Nationwide distribution in the states of PA, NC, CO, CA, LA, KY, NY, AZ, MA, LA, OK.
Product description
Ion Torrent Genexus Dx with software, Catalog: A40003539, which supports the Ion Torrent Genexus Dx Integrated Sequencer
Code information (models, lots, serial numbers)
UDI-DI: 10190302018883, Software: 6.8.2.1 and earlier. Sequencer UDI-DI: 10190302017794. Sequencer Serial Numbers: 2619124040913, 2619726010247, 2619124060941, 2619124060942, 2619725090229, 2619725100238, 2619122070666, 2619725080220, 2619725090226, 2619726010251, 2619726020265, 2619725090225, 2619725100235, 2619725050205, 2619726010249, 2619122070658
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Ion Torrent™ Genexus™ Dx Integrated Sequencer A53256Life Technologies Holdings Pte. Ltd. · High Throughput Dna Sequence Analyzer1 recall0 reports
- Ion Torrent Genexus Dx Software 6.8Life Technologies Holdings Pte. Ltd. · High Throughput Dna Sequence Analyzer1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Nov 27, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Nov 27, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Nov 27, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page