Convenience Kits containing CONMED AirSeal Access Ports recallZ-0183-2027 · Medline Industries, LP · initiated Aug 31, 2026
Open, Classified All recalls posted in October 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Kits contain CONMED AirSeal Access Ports. The ports were recalled by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending procedures.
FDA root cause category: Process change control
What the company is doing
The firm began notifying customers on August 31, 2026. Customers were provided with CONMED's original recall notification, which instructs users to have a replacement AirSeal disposable be available prior to the procedure in order to minimize potential delays. Users should ensure tubing is properly connected between the tube set and access ports prior to initiating insufflation. If the connection cannot be securely tightened and the "overpressure" alert cannot be resolved, users should replace the affected device. Users were also provided with instructions on how to connect the AirSeal tubeset and access ports, which include new figures as an update to the prior Instructions for Use. The corrected Instructions for Use are available electronically at eifu.conmed.com/eifu.
Recall details
- Recall number
- Z-0183-2027
- Recall class
- Not yet classified by the FDA
- Status
- Open, Classified
- Initiated by the company
- Aug 31, 2026
- Posted by the FDA
- Oct 15, 2026
- Recalling firm
- Medline Industries, LP
- Quantity
- 3,551 kits
- Product code
- OHD
- Distribution
- US Nationwide distribution in the states of AZ, CA, IN, MD, MO, NJ, OH, and WI.
Product description
Convenience Kits containing CONMED AirSeal Access Ports CHN GYN ONCOLOGY DYNJ906731K DYNJ906731L DYNJ906731N DAVINCI PROSTATECTOMY DYNJG901002A DYNJG901002B DYNJG901002C DYNJG901002D GENERAL DAVINCI DYNJG901004B DYNJG901004D HERNANDEZ DAVINCI ACC BUNDLE M DYNJ911258A KIT ROBOTICS UROLOGY PROSTATE DYKMBNDL116D DYKMBNDL116F ROBOTIC GYN DYNJ907827F ROBOTIC PACK DYNJ50474G RR-DAVINCI PROSTATE ACCESSORY DYNJ63392C DYNJ63392F
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
DYNJ906731K UDI-DI 10195327436346 Lots 24ADB566 24CDB291 24CDB969 24GDB449 24JDA590 24KDB303 25BDB494 25CDA446 25EDB457 25FDA837 25FDB582 25GDA082 DYNJ906731L UDI-DI 10198459403675 Lots 25HDB211 25JDA226 DYNJ906731N UDI-DI 10198459758249 Lots 26EDA043 26FDB001 26FDB487 DYNJG901002A UDI-DI 10195327401481 Lot 24BLA300 DYNJG901002B UDI-DI 10195327677145 Lots 24GLA290 24ILA519 24JLA479 24LLA939 25ALA154 25ELA210 25FLA112 25GLA399 25HLA560 DYNJG901002C UDI-DI 10198459545672 Lots 25JLA619 25KLA897 26ALA207 26BLA625 26CLA984 DYNJG901002D UDI-DI 10198459723742 Lots 26ELA161 26ELA744 26GLA765 26HLA179 DYNJG901004B UDI-DI 10195327677138 Lots 24FLA735 24GLA836 24KLA502 25ALA409 25ALA837 25DLA322 25ELA451 25FLA291 25GLB051 25HLA207 25ILA840 25JLA027 25JLA621 DYNJG901004D UDI-DI 10198459723759 Lots 26DLA382 26ELA743 26FLB123 DYNJ911258A UDI-DI 10198459582097 Lot 26AMF842 DYKMBNDL116D UDI-DI 10195327608866 Lots 24CBF889 24DBA349 DYKMBNDL116F UDI-DI 10198459000706 Lots 24DBS388 24FBC411 24GBK948 24IBV184 24JBB328 DYNJ907827F UDI-DI 10195327345365 Lots 24ABV361 24CBM651 24DBD866 24GBH678 DYNJ50474G UDI-DI 10198459140235 Lots 24KDA630 25ADB023 25BDB826 25CDB231 25DDA066 25HDA254 25JDA520 25LDA008 25LDC029 26DDA900 26FDA459 26FDB573 DYNJ63392C UDI-DI 10195327634643 Lots 24BLA318 24CLA990 24ELA575 24FLA977 24HLA137 24ILA935 DYNJ63392F UDI-DI 10198459315572 Lots 25DLA944 25ELA626 25FLA272 25GLA905 25HLA240 25HLA894 25JLA091 25KLA775 25LLA985 26ALA670 26DLB121 26FLA701
Devices named in this recall
10 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYKMBNDL116DMedline Industries, Inc. · Gynecological laparoscopic kit3 recalls0 reports
- Medline Industries, Inc. DYKMBNDL116FMedline Industries, Inc. · Gynecological laparoscopic kit3 recalls0 reports
- Medline Industries, Inc. DYNJ63392FMedline Industries, Inc. · Gynecological laparoscopic kit3 recalls0 reports
- Medline Industries, Inc. DYNJ906731KMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJG901002BMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ907827FMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJG901002AMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ63392CMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJG901004BMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ50474GMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99888).
- Z-0180-2027 · Open, ClassifiedConvenience Kits containing CONMED AirSeal Access Ports ADVANCE LAPAROSCOPIC DYNJ911135 LAP CHOLE DYNJ911660 LAP NISSEN DYNJ911560 UROLOGY LAPAROSCOPIC - 51638 DYNJ90…
- Z-0181-2027 · Open, ClassifiedConvenience Kits containing CONMED AirSeal Access Ports HERNANDEZ DAVINCI ACC BUNDLE M DYNJ911258 KIT ROBOTICS UROLOGY PROSTATE DYKMBNDL116I ROBOTIC CHEST DYNJ910997 DYNJ…
- Z-0182-2027 · Open, ClassifiedConvenience Kits containing CONMED AirSeal Access Ports GENERAL DAVINCI DYNJG901004C LAP CHOLE DYNJ88071 ROBOTIC PACK DYNJ50474F
- Z-0185-2027 · Open, ClassifiedConvenience Kits containing CONMED AirSeal Access Ports KIT ROBOTICS UROLOGY PROSTATE DYKMBNDL116G
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 21, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 21, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 21, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page